LeadCare® II Replacement AC Adapter and Plug Set
by Magellan Diagnostics, Inc. United States GUDID
Product code DOF
Regulatory status
Recall ongoing. 6 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00850355006123 | Date19 Sep 2016 | StatusIn distribution |
Coverage
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-0162-2027
In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results.
-
- Ongoing Z-1895-2026
- Ongoing Z-1893-2026
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
-
- Ongoing Z-0584-2026
Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
-
- Ongoing Z-0219-2024
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
-
- Ongoing Z-1597-2022
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
-
- Terminated Z-1952-2021
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
-
- Terminated Z-0285-2019
Defect with the calibration button for lot 1808M. Level 1 and Level 2 Quality Control (QC) values will generate results of HIGH (>65 g/dL). Results for patient testing with this lot will be displayed as HIGH.
-
- Terminated Z-1295-2018
The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
-
- Terminated Z-2226-2017
Underestimates the lead concentration of venous blood samples when the sample is analyzed with the LeadCare systems
-
- Terminated Z-3035-2017
LeadCare II Test Kits controls out of range
-
- Terminated Z-0042-2008
- Terminated Z-0043-2008
Level 1 and Level 2 Control values were incorrectly assigned.
Specification
- Generic Name
- Lead (Pb) analyser IVD
- Model / Ref / Catalog
- LeadCare® II, 82-0001
- Primary DI / UDI-DI
- 00850355006123
- 510(k) Number
- FDA Product Code
- DOF
- GMDN Term
- Lead (Pb) analyser IVD
- Device Class
- Class II
- Regulation Number
- 862.3550
- Commercial Distribution
- In Commercial Distribution
Properties & raw data
Show properties, registry sources and raw records
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No
Sources (1)
- FDA UDI 16747fe1-c01a-46d1-ad1a-9a106ee7b2c0