LeadCare® Plus Blood Lead Analyzer Kit
by Magellan Diagnostics, Inc. United States GUDID
Product code DOF
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00850355006062 | Date19 Sep 2016 | StatusIn distribution |
Coverage
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-1894-2026
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
-
- Terminated Z-1953-2021
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
-
- Terminated Z-1296-2018
The incorrect placement of AA batteries into certain LeadCare Plus Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
-
- Terminated Z-1926-2017
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
Specification
- Generic Name
- Lead (Pb) analyser IVD
- Model / Ref / Catalog
- LeadCare® Plus, 82-0001
- Primary DI / UDI-DI
- 00850355006062
- 510(k) Number
- FDA Product Code
- DOF
- GMDN Term
- Lead (Pb) analyser IVD
- Device Class
- Class II
- Regulation Number
- 862.3550
- Commercial Distribution
- In Commercial Distribution
Properties & raw data
Show properties, registry sources and raw records
Other properties (11)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No
Sources (1)