Leksell Gammaplan
by Patterson Consulting Group, Inc.
510(k) K973441
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK973441 | Date1 Apr 1998 | StatusActive |
Recalls
-
- Terminated Z-0307-2012
There is a risk that the user may interpret the instructions in the manual too literally, and adjust the TMR 10 dose description to obtain an ETCPD that matches previous clinical prescriptions, which in turn could lead to a prescription dose that deviates a few percent from the intended dose.
-
- Terminated Z-1134-2010
Although the co-registration looks good during the verification step in the co-registration dialog, the obtain transformation may include an error that depends on the voxel sizes and acquisition parameters of the co-registered image studies.
-
- Terminated Z-1604-2010
The precision of the calculation used to create the fused study in LGP 8.0 is too low and should not be used until the system is upgraded to LGP 8.2.
-
- Terminated Z-1223-2011
Investigation found that if a user accidentally selects the wrong image/tube position images can be displayed flipped.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5750
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent