Leucadia Autolok Pedicle Screw System
by Phygen, LLC
510(k) K113366
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK113366 | Date17 Jan 2012 | StatusActive |
Recalls
-
- Terminated Z-1421-2013
Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia" Autolok set screw may interfere with the top of the screw head prior to fully locking the rod. If this condition occurs, the polyaxial screw head and rod may fail to provide a rigid construct under loading. To date there have been no reports of patient injury, adve
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- NKB
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3070
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent