Level Sensor Pad
by Maquet Cardiopulmonary GmbH
Ref 701022208
Level Sensor Pad (701022208) is a cardiopulmonary bypass system (EU MDD Class IIb; FDA Class II; FDA product code DTQ, Console, Heart-Lung Machine, Cardiopulmonary Bypass) made by Maquet Cardiopulmonary GmbH, registered in EUDAMED on 10 Apr 2026 (Basic UDI-DI B-04037691687308), cleared under FDA 510(k) K102726 on 4 Apr 2011 and listed in FDA GUDID since 8 Dec 2020 (UDI-DI 04037691687308), active as of 22 Oct 2026. 8 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
3 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDD EUDAMED | NumberB-04037691687308 04037691687308 | Date10 Apr 2026 | StatusActive |
| United States | GUDID listing GUDID | Number04037691687308 | Date8 Dec 2020 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK102726 | Date4 Apr 2011 | StatusActive |
Official records
- EUDAMED public site B-04037691687308
- AccessGUDID record 04037691687308
- FDA 510(k) database K102726
From the FDA record
- Device description
- Level Sensor Pads LSP 20-531, non-sterile, box of 100 pieces
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI B-04037691687308 registered 10 Apr 2026 Active
- FDA 510(k) K102726 decided 4 Apr 2011 Active
- FDA GUDID UDI-DI 04037691687308 listed 8 Dec 2020 Listed
Recalls
-
- Ongoing Z-1823-2025
HKH 8820 Wall Holder does not comply with standard DIN EN 1789:20 and has the potential to detach.
-
- Ongoing Z-1662-2023
Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.
-
- Ongoing Z-1643-2023
1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.
-
- Terminated Z-0986-2019
Upon servicing the gearwheel in the Emergency Drive the gearwheel may have been re-assembled incorrectly, leading to the inadequate operation of the Emergency Drive.
-
- Terminated Z-1642-2018
There is a possibility for the clamps of the gas bottle holder to crack or break when exposed to strong vibrations.
-
- Terminated Z-1372-2016
Maquet Cardiopulmonary (MCP) is conducting a voluntary recall based on the potential risk of exposure to endotoxin for various products.
-
- Terminated Z-1690-2015
A potential issue was detected that could result in blood leakage from the gas outlet port in all lots of Quadrox-iD Adult Diffusion Membrane Oxygenator with Bioline and Softline Coatings, Custom Tubing Packs and Diffusion Membrane HLS Set Advanced and HIT Set Advanced Models.
-
- Terminated Z-0211-2015
Cardiohelp- i System may have a software issue that can potentially result in an erroneous display of a "Battery Needs Service" message after startup of the units when using either AC or DC power
-
- Terminated Z-0082-2014
It has come to the attention of MAQUET that in certain rare instances, the CARDIOHELP-i device may display an error message with audible alarm, indicating "battery defective." The alarm will occur-upon start up or when the main power supply is unplugged from the main power source. The human machine interface touch screen of the CARDIOHELP-i device may gradually become black due to the human machine interface (HMI) back light shutting down.
-
- Terminated Z-1149-2013
It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touch screen) of the CARDIOHELP device. The human machine interface (HMI) is the central display of the CARDIOHELP device, where measured values, settings and alarms are displayed and adjusted/confirmed, with the exception of bl
Timeline
- Recall initiated FDA recalls Z-0211-2015
- Recall initiated FDA recalls Z-1690-2015
- Recall initiated FDA recalls Z-1372-2016
- Recall initiated FDA recalls Z-1642-2018
- Recall initiated FDA recalls Z-0986-2019
- Listed FDA GUDID 04037691687308
- Recall initiated FDA recalls Z-1662-2023
- Recall initiated FDA recalls Z-1643-2023
- Recall initiated FDA recalls Z-1823-2025
- Registered EUDAMED B-04037691687308
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Level Sensor Pad EUDAMEDLevel Sensor Pads GUDID
- Generic name
- Cardiopulmonary bypass system, roller type
- Model / Ref / Catalog
- 701022208, LSP 20-531
- Primary DI / UDI-DI
- 04037691687308
- Basic UDI-DI
- B-04037691687308
- 510(k) Number
- FDA product code
- DTQ
- EMDN code
- Z12050280 Extra-corporeal circulation instruments - hardware accessories
- GMDN term
- Cardiopulmonary bypass system, roller type
- Device class
- Class IIb EUDAMEDClass II GUDID
- Regulation number
- 870.4220
- EU registration type
- Legacy device (under a directive)
- EU market of registration
- Germany
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / reference | 701022208 | LSP 20-531 |
| Device class | Class IIb | Class II |
Manufacturer details
- EUDAMED SRN
- DE-MF-000020091
Similar devices in this category
Other devices filed under EXTRA-CORPOREAL CIRCULATION INSTRUMENTS.
- Quantum 52-000027-00 by SPECTRUM MEDICAL LIMITED
- Quantum 52-000096-00 by SPECTRUM MEDICAL LIMITED
- MAQUET CARDIOHELP FBS 3/8" x 3/32" L0.9 70104.8001 by Maquet Cardiopulmonary AG
- MAQUET CARDIOHELP Venous Probe 70104.8803 by Maquet Cardiopulmonary AG
- MPS Water Hose Kit 5303030 by QUEST MEDICAL INC
- MPS Stand Alone Cart 5001005 by QUEST MEDICAL INC
- ESSENZ Patient Monitor Software by LivaNova Deutschland GmbH
- Electronic Gas Blender 2 l/min by LivaNova Deutschland GmbH
- Solution Delivery Cassettes 1467 by TransMedics, Inc
- OCS Lung Console 2000 by TransMedics, Inc