Liat (Tm) Influenza a / B Assay, Liat (Tm) Analyzer
by Iquum, Inc.
510(k) K111387
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK111387 | Date4 Aug 2011 | StatusActive |
Recalls
-
- Ongoing Z-0828-2022
Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influenza B results.
-
- Terminated Z-2203-2016
Kit labeled with the incorrect expiration date.
-
- Terminated Z-2216-2016
Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results.
-
- Terminated Z-2543-2015
Due to a supplier issue, the incorrect green and amber LED components were used in the manufacture of the LED printed circuit board assemblies (PCBA) in certain cobas Liat Analyzers.