Licox® BC10
by Integra Lifesciences (Ireland) Limited
Part of a family of 24 devices: see the family page
Ref BC10
Licox® BC10 is a device classified as Class I in the EU and Class II in the US made by Integra Lifesciences (Ireland) Limited, registered in EUDAMED on 8 Jul 2026 (Basic UDI-DI 038178B0036CSS3D), cleared under FDA 510(k) K002765 on 24 Nov 2000 and listed in FDA GUDID since 2 Aug 2016 (UDI-DI 10381780250128), active as of 23 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
None ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number038178B0036CSS3D 10381780250128 | Date8 Jul 2026 | StatusActive |
| United States | GUDID listing GUDID | Number10381780250128 | Date2 Aug 2016 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK002765 | Date24 Nov 2000 | StatusActive |
Official records
- EUDAMED public site 038178B0036CSS3D
- AccessGUDID record 10381780250128
- FDA 510(k) database K002765
Across registries
- EUDAMED Basic UDI-DI 038178B0036CSS3D registered 8 Jul 2026 Active
- FDA 510(k) K002765 decided 24 Nov 2000 Active
- FDA GUDID UDI-DI 10381780250128 listed 2 Aug 2016 Listed
Recalls
-
- Terminated Z-0831-2015
Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.
-
- Terminated Z-1518-2011
Certain LICOX CMP Tissue Oxygen Monitors could allow for underestimation of temperature values.
-
- Terminated Z-0622-05
LICOX Model C8.B temperature probes manufactured by GMS mbH, an Integra company, may exhibit progen (endotoxin) levels above the FDA limit of 2.15 endotoxin units per device. Elevated endotoxin levels may result in an increased patient risk of fever.
Timeline
- Cleared FDA 510(k) K002765
- Recall initiated FDA recalls Z-0622-05
- Recall initiated FDA recalls Z-1518-2011
- Recall initiated FDA recalls Z-0831-2015
- Listed FDA GUDID 10381780250128
- Registered EUDAMED 038178B0036CSS3D
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Licox®
- Generic name
- Electrical-only medical device connection cable, reusable
- Device name
- lntegra Monitor cables
- Model / Ref / Catalog
- BC10
- Primary DI / UDI-DI
- 10381780250128
- Basic UDI-DI
- 038178B0036CSS3D
- 510(k) Number
- FDA product code
- GWM
- EMDN code
- L99 Surgical instruments, reusable - other
- GMDN term
- Electrical-only medical device connection cable, reusable
- Device class
- Class I EUDAMEDClass II GUDID
- Regulation number
- 882.1620
- EU registration type
- Standard device
- EU market of registration
- Belgium
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Unsafe
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class I | Class II |
Similar devices in this category