LifePak 12 Defibrillator/Monitor
by Medtronic Emergency Response Systems, Inc. 2 variants
510(k) K063510
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK063510 | Date26 Jan 2007 | StatusActive |
Recalls
-
- Terminated Z-0811-2010
Affected devices contain a Biphasic PCBA that may have a solder defect on the H Bridge component that could prevent the device from providing a defibrillation shock. There could be a delay in therapy or prevention of defibrillation therapy if the defibrillator were needed in a cardiac arrest situation. Delay or prevention of defibrillation therapy may contribute to the death of a viable patient.
-
- Terminated Z-1904-2008
The LIFEPAK 12 defibrillator / monitors with software version 130 have an increase in likelihood for an incorrect Shock Advisory Algorithm (SAS) decision if the Auto Analyze setting in AED mode is On, leading to incorrectly render of shock or no shock decision. When Auto Analyze is set to On in AED mode the device initiates the SAS analysis immediately (no waiting period or warning prior to analys
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MKJ
- Device Class
- Class III
- Regulation
- 21 CFR Part 870.5310
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 18 Sep 2026
The LifePak 12 Defibrillator/Monitor is an automated external defibrillator (non-wearable) designed to deliver synchronized and unsynchronized cardioversion and defibrillation. It provides real-time ECG monitoring, rhythm analysis, and automated shock advisory for emergency cardiac care, supporting both manual and semiautomatic operation.
Supplied as a reusable medical device, the LifePak 12 is intended for use in acute cardiac care settings, including hospital and pre-hospital environments. It is regulated under FDA 510(k) clearance (K063510) as a Class III device, adhering to 21 CFR 870.5310. The device is not MRI-compatible and requires proper maintenance, calibration, and storage as outlined in manufacturer guidelines.