  "format": "devicerts.device.v1",
  "cite_as": "LifePak 500 Biphasic, Physio-Control Corp. deviCERTS medical device catalogue. https://devicerts.com/devices/lifepak-500-biphasic-mkj (data as of 2026-11-23).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/lifepak-500-biphasic-mkj",
    "json_url": "https://devicerts.com/devices/lifepak-500-biphasic-mkj.json",
    "markdown_url": "https://devicerts.com/devices/lifepak-500-biphasic-mkj.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-23",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/lifepak-500-biphasic-mkj",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "LifePak 500 Biphasic",
    "registry_name": "LIFEPAK 500 BIPHASIC",
    "url": "https://devicerts.com/devices/lifepak-500-biphasic-mkj",
    "manufacturer": {
      "name": "Physio-Control Corp.",
      "url": "https://devicerts.com/manufacturers/physio-control-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class III",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K983393",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "MKJ"
    ],
    "emdn": [],
    "gmdn": [],
    "description": {
      "text": "The LifePak 500 Biphasic is an Automated External Defibrillator (Non-Wearable), designed to deliver biphasic waveform shocks for the resuscitation of patients experiencing sudden cardiac arrest. It provides automated analysis of heart rhythms and delivers appropriate defibrillation therapy when indicated, with voice prompts to guide rescuers through the resuscitation process.\n\nSupplied as a reusable device, the LifePak 500 Biphasic is intended for use in pre-hospital and in-hospital settings, including emergency medical services, hospitals, and public access defibrillation programs. It is classified as a Class III medical device under FDA regulations (870.5310) and holds a 510(k) clearance with tracking requirements. The device is not MRI-safe and requires proper storage and maintenance as specified by the manufacturer.",
      "ai_generated": true,
      "generated_on": "2026-09-18",
      "note": "Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data is the record."
    },
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K983393",
      "number_label": null,
      "device_udi_di": null,
      "date": "1999-05-05",
      "date_kind": "decision",
      "decision_code": "ST",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1633-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-03-20",
      "reason": "Automated External Defibrillators  may not detect a patient connection due to mechanical wear-through of the contact plating on the therapy connector. The wear-through could result in the device not recognizing a patient is connected which could result in a delay in treatment",
      "applies": "this device",
      "via": "K983393"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K983393",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "MKJ",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class III",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 870.5310",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Cardiovascular",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "ST",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}