  "format": "devicerts.device.v1",
  "cite_as": "Lifepak 500, Physio-Control Corp. deviCERTS medical device catalogue. https://devicerts.com/devices/lifepak-500-mkj (data as of 2026-11-19).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/lifepak-500-mkj",
    "json_url": "https://devicerts.com/devices/lifepak-500-mkj.json",
    "markdown_url": "https://devicerts.com/devices/lifepak-500-mkj.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-19",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/lifepak-500-mkj",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Lifepak 500",
    "registry_name": "LIFEPAK 500",
    "url": "https://devicerts.com/devices/lifepak-500-mkj",
    "manufacturer": {
      "name": "Physio-Control Corp.",
      "url": "https://devicerts.com/manufacturers/physio-control-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class III",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K955854",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "MKJ"
    ],
    "emdn": [],
    "gmdn": [],
    "description": {
      "text": "The LifePak 500 is an Automated External Defibrillator (Non-Wearable), designed to deliver biphasic defibrillation waveforms for the treatment of cardiac arrest due to ventricular fibrillation or pulseless ventricular tachycardia. It provides automated analysis of heart rhythms and delivers synchronized cardioversion or defibrillation shocks as clinically indicated, with adaptive biphasic technology to optimize energy delivery.\n\nThis device is intended for use in emergency cardiac resuscitation by trained responders, including healthcare professionals, first responders, and lay rescuers. Supplied as a reusable medical device, it requires proper maintenance and calibration per manufacturer guidelines. The LifePak 500 is classified as a Class III device under FDA regulations (870.5310) and holds a Traditional 510(k) clearance with a Cardiovascular advisory committee review. It is not MRI-compatible and must be stored in accordance with manufacturer instructions to ensure operational readiness.",
      "ai_generated": true,
      "generated_on": "2026-09-18",
      "note": "Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data is the record."
    },
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K955854",
      "number_label": null,
      "device_udi_di": null,
      "date": "1996-11-04",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1633-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-03-20",
      "reason": "Automated External Defibrillators  may not detect a patient connection due to mechanical wear-through of the contact plating on the therapy connector. The wear-through could result in the device not recognizing a patient is connected which could result in a delay in treatment",
      "applies": "this device",
      "via": "K955854"
    },
    {
      "number": "Z-1148-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2007-06-13",
      "reason": "Reduced Shock- the device may deliver 100 Joule (J) to a patient, which is less than the recommended factory minimum default setting of 200 J.",
      "applies": "shared submission",
      "via": "K955854"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K955854",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "MKJ",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class III",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 870.5310",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Cardiovascular",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}