LifeSPARC 5800-1000
by Cardiacassist, Inc. GUDID
Withdrawn
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00814112020609 | Date29 May 2020 | StatusWithdrawn |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-2465-2023
Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,
-
- Terminated Z-0926-2023
A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The Critical Failure was first addressed in the firm's recall initiated July 21, 2022. During the Critical Failure, the software freezes or crashes and the screen does not display data.
-
- Terminated Z-1763-2022
Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.
Specification
- Trade Name
- LifeSPARC
- Generic Name
- Cardiopulmonary bypass system centrifugal pump
- Model / Reference
- 5800-1000
- Primary DI / UDI-DI
- 00814112020609
- FDA Product Code
- KRI, KFM, DTZ
- GMDN Term
- Cardiopulmonary bypass system centrifugal pump
- Device Class
- Class I
- Regulation Number
- 870.4200, 870.4360, 870.4350
- Commercial Distribution
- Not in Commercial Distribution
- MRI Safety
- MR Unsafe
Other devices from Cardiacassist, Inc.
Properties
Other properties (7)
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No