Light Guide Adapter MAJ-1413
by Olympus Medical Systems Corporation
Light Guide Adapter MAJ-1413 (N1773933) is a light source adaptor (FDA Class II; FDA product code FAJ, Cystoscope And Accessories, Flexible/Rigid) made by Olympus Medical Systems Corporation, cleared under FDA 510(k) K221690 on 29 Dec 2022 and listed in FDA GUDID since 22 Apr 2023 (UDI-DI 04953170434488), active as of 22 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number04953170434488 | Date22 Apr 2023 | StatusListed |
| United States | FDA 510(k)SESE | NumberK221690 | Date29 Dec 2022 | StatusActive |
Official records
- AccessGUDID record 04953170434488
- FDA 510(k) database K221690
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use
Across registries
- FDA 510(k) K221690 decided 29 Dec 2022 Active
- FDA GUDID UDI-DI 04953170434488 listed 22 Apr 2023 Listed
Recalls
-
- Ongoing Z-0685-2024
- Ongoing Z-0684-2024
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
- Ongoing Z-0331-2024
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Timeline
- Cleared FDA 510(k) K221690
- Listed FDA GUDID 04953170434488
- Recall initiated FDA recalls Z-0331-2024
- Recall initiated FDA recalls Z-0685-2024
- Recall initiated FDA recalls Z-0684-2024
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Light Guide Adapter
- Generic name
- Light source adaptor
- Model / Ref / Catalog
- MAJ-1413, N1773933
- Primary DI / UDI-DI
- 04953170434488
- FDA product code
- FAJ
- GMDN term
- Light source adaptor
- Device class
- Class II
- Regulation number
- 876.1500
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category