Lightspeed 7.0 CT Scanner System
510(k) K040372
Lightspeed 7.0 CT Scanner System is a medical device (FDA Class II; FDA product code JAK, System, X-Ray, Tomography, Computed) made by GE Medical Systems, LLC, cleared under FDA 510(k) K040372 on 1 Mar 2004, active as of 22 Oct 2026. 5 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK040372 | Date1 Mar 2004 | StatusActive |
Official records
- FDA 510(k) database K040372
From the FDA record
- Received by the FDA
- 17 Feb 2004
- FDA decision
- 1 Mar 2004 (Substantially Equivalent)
- FDA review time
- 13 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-2183-2017
It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.
-
- Terminated Z-1188-2013
Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil conduit breaking, spilling oil within the gantry and outside the system. This issue affects GE Healthcare VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspe
-
- Terminated Z-0803-2007
- Terminated Z-0802-2007
Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.
-
- Terminated Z-0507-2007
When the manual cradle release feature on the table is used to move a patient out of the gantry, a discrepancy of up to 6mm between the numerical display on the gantry and the actual position of the table could be introduced.
Timeline
- Cleared FDA 510(k) K040372
- Recall initiated FDA recalls Z-0507-2007
- Recall initiated FDA recalls Z-0803-2007
- Recall initiated FDA recalls Z-0802-2007
- Recall initiated FDA recalls Z-1188-2013
- Recall initiated FDA recalls Z-2183-2017
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- JAK
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1750
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).
- ITERATIVE RECONSTRUCTION TECHNIQUE SOFTWARE APPLICATION by Philips Healthcare (Cleveland)
- CardIQ Xpress 2.0 with Snapshot Freeze Option CardIQ Xpress 2.0 Reveal by GE Medical Systems SCS
- Al-Rad Companion (Pulmonary) by SIEMENS MEDICAL SOLUTIONS USA, INC.
- MYOPERFUSION by Toshiba Medical Systems
- SOMATOM SMILE by SIEMENS MEDICAL SOLUTIONS USA, INC.
- SCAN BAND by Cook, Inc.
- HEAD-SPINE STABILIZER HSS-1 by Imaging Accessories, Inc.
- ADVANCE VESSEL ANALYSIS II by GE Healthcare
- VIATRONIC V3D CALCIUM SCORING by Viatronix, Inc.
- PRIMELUNG by Accuimage Diagnostics Corp.