  "format": "devicerts.device.v1",
  "cite_as": "Lightspeed 7.0 CT Scanner System, GE Medical Systems, LLC. deviCERTS medical device catalogue. https://devicerts.com/devices/lightspeed-7-0-ct-scanner-system-jak (data as of 2026-11-22).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/lightspeed-7-0-ct-scanner-system-jak",
    "json_url": "https://devicerts.com/devices/lightspeed-7-0-ct-scanner-system-jak.json",
    "markdown_url": "https://devicerts.com/devices/lightspeed-7-0-ct-scanner-system-jak.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-22",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/lightspeed-7-0-ct-scanner-system-jak",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Lightspeed 7.0 CT Scanner System",
    "registry_name": "LIGHTSPEED 7.0 CT SCANNER SYSTEM",
    "url": "https://devicerts.com/devices/lightspeed-7-0-ct-scanner-system-jak",
    "manufacturer": {
      "name": "GE Medical Systems, LLC",
      "url": "https://devicerts.com/manufacturers/ge-medical-systems-llc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K040372",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "JAK"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K040372",
      "number_label": null,
      "device_udi_di": null,
      "date": "2004-03-01",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2183-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-05-19",
      "reason": "It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.",
      "applies": "this device",
      "via": "K040372"
    },
    {
      "number": "Z-1188-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2011-07-25",
      "reason": "Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil conduit breaking, spilling oil within the gantry and outside the system. This issue affects GE Healthcare VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspe",
      "applies": "this device",
      "via": "K040372"
    },
    {
      "number": "Z-0803-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2006-12-21",
      "reason": "Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.",
      "applies": "this device",
      "via": "K040372"
    },
    {
      "number": "Z-0802-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2006-12-21",
      "reason": "Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.",
      "applies": "this device",
      "via": "K040372"
    },
    {
      "number": "Z-0507-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2006-12-08",
      "reason": "When the manual cradle release feature on the table is used to move a patient out of the gantry, a discrepancy of up to 6mm between the numerical display on the gantry and the actual position of the table could be introduced.",
      "applies": "this device",
      "via": "K040372"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K040372",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "JAK",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1750",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}