Skip to content

Lipase Flex Reagent Cartridge

by Dupont Medical Products

510(k) K952816

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

3 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK952816 Date20 Jul 1995ClearedStatusActive

Recalls

3 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-0466-2019 FDA Class II initiated 7 Sep 2018
    • Terminated Z-0465-2019 FDA Class II initiated 7 Sep 2018

    There is lot to lot variability for QC and patient results when using the identified reagent cartridge lots.

    • Terminated Z-0588-2011 FDA Unclassified (pre-2012) initiated 10 Mar 2010

    Open Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
CHI
Device Class
Class I
Regulation
21 CFR Part 862.1465
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent

Other devices from Dupont Medical Products

See all devices from Dupont Medical Products →