Lipase Flex Reagent Cartridge
510(k) K952816
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK952816 | Date20 Jul 1995 | StatusActive |
Recalls
-
- Terminated Z-0466-2019
- Terminated Z-0465-2019
There is lot to lot variability for QC and patient results when using the identified reagent cartridge lots.
-
- Terminated Z-0588-2011
Open Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CHI
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1465
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent