Low Plasticizer Administration Sets
by Travenol Laboratories, S.A.
510(k) K791496
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK791496 | Date8 Apr 1980 | StatusActive |
Recalls
-
- Terminated Z-0629-06
- Terminated Z-0626-06
- Terminated Z-1424-05
The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
-
- Terminated Z-1103-05
Leaks and disconnection between the bifurcated Y-site and the set tubing may occur and result in underdelivery of medication, blood loss, and a breach of the sterile fluid pathway.
-
- Terminated Z-0854-05
Inadequate directions for use resulting in fluid leaks, valves sticking in the open position, and flow output restriction.
-
- Terminated Z-0845-05
Possible occlusion of the Clearlink injection site valve due to inadequate silicone lubrication of the slit in the upper seal of the valve.
-
- Terminated Z-0694-05
Product older than three years may not operate as intended.
-
- Terminated Z-0519-03
- Terminated Z-0520-03
- Terminated Z-0521-03
- Terminated Z-0517-03
- Terminated Z-0518-03
The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FPA
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5440
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent