Ltv 1000
510(k) K984056
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK984056 | Date12 Apr 1999 | StatusActive |
Recalls
-
- Terminated Z-1464-2008
- Terminated Z-1469-2008
- Terminated Z-1466-2008
- Terminated Z-1470-2008
- Terminated Z-1465-2008
- Terminated Z-1467-2008
- Terminated Z-1468-2008
- Terminated Z-1463-2008
The failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
-
- Terminated Z-0185-06
- Terminated Z-0186-06
- Terminated Z-0184-06
Pulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe for the patient. This malfunction may occur without an accompanying audible alarm.
-
- Terminated Z-1485-04
Ventilators have not consistently switched to internal battery operation when the external DC power source has become inadequate to supply stable power. This situation results in loss of ventilation to the patient.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CBK
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5895
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent