Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom Rax
by Siemens Medi Cal Solutions, Inc.
510(k) K173639
Active
Valid until
No expiry
FDA recalls
13 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK173639 | Date2 Apr 2018 | StatusActive |
Recalls
-
- Ongoing Z-0999-2026
A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ programs (OGPs) the collimation may be incorrect.
-
- Ongoing Z-1982-2025
- Ongoing Z-1981-2025
A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.
-
- Ongoing Z-2527-2024
- Ongoing Z-2525-2024
- Ongoing Z-2526-2024
The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
-
- Ongoing Z-1167-2023
- Ongoing Z-1168-2023
- Ongoing Z-1165-2023
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
-
- Ongoing Z-1002-2023
There is potential for the footrest to detach from the patient table during use. Device intended to visualize anatomical structures by converting an Xray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract. The unit may also be used in emergency applications, lymphography, endoscopy, myelography, venography, arthrography, interventional radiology, digital angiography, and digital subtraction angiography (DSA). The system may be used on pediatric, adult, and bariatric patients. These devices are not for mammography examinations.
-
- Ongoing Z-1480-2023
- Ongoing Z-1481-2023
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.
-
- Ongoing Z-1479-2023
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube. Siemens Healthineers became aware of one customer site where this issue was identified during clinical operation.
-
- Terminated Z-0806-2021
After the March 2019 corrective action on the monitor support arm, there is a residual risk of the support arm becoming dislodged in cases when the monitor support arm is used intensely over a long period of time.
-
- Terminated Z-1332-2020
- Terminated Z-1333-2020
- Terminated Z-1331-2020
Siemens became aware of an issue of incorrect use of the bypass key overriding the built-in collision control creating risk of collision with obstacles or persons when the operator uses the override function.
-
- Terminated Z-1338-2020
- Terminated Z-1337-2020
A software bug may lead to one image to be assigned to two different patients, which could potentially affect medical diagnosis.
-
- Terminated Z-2756-2020
- Terminated Z-2755-2020
- Terminated Z-2757-2020
Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis
-
- Terminated Z-0985-2019
The swivel arm holding the monitors may become dislodged from the carrying arm.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent