Skip to content

LUX-Dx II (M302); LUX-Dx II+ (M312)

by Boston Scientific

510(k) K231328

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

6 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 9 Oct 2026. What this means

Regulatory status

as of registry sync 9 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK231328 Date19 Aug 2023ClearedStatusActive

Recalls

6 recalls · 6 ongoing · FDA

  • 3 recalls, same reason

    • Ongoing Z-3081-2026 FDA Class II initiated 6 Aug 2026
    • Ongoing Z-3083-2026 FDA Class II initiated 6 Aug 2026
    • Ongoing Z-3082-2026 FDA Class II initiated 6 Aug 2026

    There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.

  • 2 recalls, same reason

    • Ongoing Z-2048-2026 FDA Class II initiated 30 Mar 2026
    • Ongoing Z-2049-2026 FDA Class II initiated 30 Mar 2026

    For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.

    • Ongoing Z-0226-2025 FDA Class II initiated 6 Sep 2024

    Patients using the Model 2939 myLUX Patient Application on their Apple mobile device (e.g., iPhone) may not be transmitting data from their Insertable Cardiac Monitor to the LATITUDE Clarity website.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
MXD
FDA applicant
Boston Scientific Corp
Device Class
Class II
Regulation
21 CFR Part 870.1025
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from Boston Scientific

See all devices from Boston Scientific →