M Series Phototherapy Equipment
by Daavlin Distributing Co. 510(k)
Active
510(k) K210881
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK210881 | Date22 Jun 2021 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-2742-2026
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Specification
- 510(k) Number
- FDA Product Code
- FTC
- Device Class
- Class II
- Regulation
- 21 CFR Part 878.4630
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent