m2000 9K15-01
Ref 9K15-01
m2000 9K15-01 is an in vitro diagnostic device (EU IVDR Class A; FDA Class II; FDA product code OOI, real time Nucleic acid amplification system) made by Abbott Molecular, registered in EUDAMED on 13 Oct 2026 (Basic UDI-DI 0884999RTK0915AL), cleared under FDA 510(k) K092705 on 28 May 2010 and listed in FDA GUDID since 1 Sep 2016 (UDI-DI 00884999031814), active as of 24 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No date given
FDA recalls
None ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/IVDR EUDAMED | Number0884999RTK0915AL 00884999031814 | Date13 Oct 2026 | StatusActive |
| United States | GUDID listing GUDID | Number00884999031814 | Date1 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK092705 | Date28 May 2010 | StatusActive |
Official records
- EUDAMED public site 0884999RTK0915AL
- AccessGUDID record 00884999031814
- FDA 510(k) database K092705
From the FDA record
- Device description
- m2000rt Instrument System
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI 0884999RTK0915AL registered 13 Oct 2026 Active
- FDA 510(k) K092705 decided 28 May 2010 Active
- FDA GUDID UDI-DI 00884999031814 listed 1 Sep 2016 Listed
Recalls
-
- Terminated Z-0463-2015
Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent Packs screen is selected.
-
- Terminated Z-2169-2014
Abbott Molecular identified that some of the 5 ml Reaction Vessels (RVs) contained in lots 56944001, 57026001, 57059001, and 57673001 (original vendor lot 2198077) may have an insufficient rim. An RV with an insufficient rim may not be held securely in the 1 mL Subsystem Carrier. This may cause the RV to be seated incorrectly in the carrier or drop through it.
-
- Terminated Z-0948-2015
The x-drive assembly may fail prematurely. This could cause the manipulator or liquid handling arm to contact an object, shake and drip, or improperly pipet.
-
- Terminated Z-0634-2013
An issue with the m2000sp Instrument software has been identified whereby, after a series of steps, if changes are made to the control or calibrator information on the Sample Extraction Assay Details screen, the changes are not saved.
-
- Terminated Z-2013-2012
In the Abbott Molecular Bar Code Scanner User's Guide, List Number 6L88-02/09, the section Modifying the Keyboard Format for Bar Code Scanner LN 6L89-01 contains incorrect bar codes that can prevent the correct configuration of the bar code scanner language format. The Abbott m2000 System is intended for use in performing nucleic acid testing in clinical laboratories. It is comprised of the Ab
-
- Terminated Z-0365-2012
Abbott Molecular has received reports in which the liquid waste sensor and components in the m2000sp lower instrument cabinet overheated and/or caught fire.
-
- Terminated Z-0048-2012
The Abbott m2000rt System software has software bug which can affect customers who manually enter a Test Order onto the m2000rt and sorts a column on the calibration or control grids of the Test Order screen by double clicking the column header. This causes a mismatch between specific concentration values and the designated calibrators and controls in the final results, which could cause an inval
Timeline
- Cleared FDA 510(k) K092705
- Recall initiated FDA recalls Z-0048-2012
- Recall initiated FDA recalls Z-0365-2012
- Recall initiated FDA recalls Z-2013-2012
- Recall initiated FDA recalls Z-0634-2013
- Recall initiated FDA recalls Z-0948-2015
- Recall initiated FDA recalls Z-2169-2014
- Recall initiated FDA recalls Z-0463-2015
- Listed FDA GUDID 00884999031814
- Registered EUDAMED 0884999RTK0915AL
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- m2000
- Generic name
- Thermal cycler nucleic acid amplification analyser IVD, laboratory
- Model / Ref / Catalog
- 9K15-01
- Primary DI / UDI-DI
- 00884999031814
- Basic UDI-DI
- 0884999RTK0915AL
- 510(k) Number
- FDA product code
- OOI
- EMDN code
- W02050116 Nucleic acid testing integrated extraction /amplification /detection systems
- GMDN term
- Thermal cycler nucleic acid amplification analyser IVD, laboratory
- Device class
- Class A EUDAMEDClass II GUDID
- Regulation number
- 862.2570
- EU registration type
- Standard device
- EU market of registration
- Germany
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class A | Class II |
Similar devices in this category
Other devices filed under NUCLEIC ACID TESTING INTEGRATED EXTRACTION /AMPLIFICATION /DETECTION SYSTEMS.
- GeneXpert® GX-INFINITY 80 by CEPHEID
- GeneXpert® GeneXpert IV (GX-IV) by CEPHEID
- Nucleic Acid Extractor BH003 by Shandong Bohong Gene Technology Co., Ltd.
- MagFast Single-Cartridge Genomic DNA Purification Kit From Blood MGF06 by Vitrosens Biotechnology Inc.
- HighPrep PCR - DX (500 mL) AC-60500E by MagBio Genomics, Inc.
- BIOFIRE ® SPOTFIRE ® Module FAST-ASY-0002 by BioFire Diagnostics, LLC
- QuantStudio™ 7 Pro Dx Real‑Time PCR Instrument A50745 by Life Technologies Holdings Pte Ltd.
- ePlex Instrument by Genmark Diagnostics, Incorporated
- Exicycler™ 96 A-2060-1 by Bioneer Corporation
- MagFast 16-Flex Genomic DNA Purification Kit From Blood MGF08 by Vitrosens Biotechnology Inc.