Skip to content

MAGEC® System

by Nuvasive Specialized Orthopedics, Incorporated

510(k) K171791

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 10 Oct 2026. What this means

Regulatory status

as of registry sync 10 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK171791 Date31 Aug 2017ClearedStatusActive

Recalls

2 recalls · 1 ongoing · FDA

    • Ongoing Z-1657-2025 FDA Class II initiated 3 Apr 2025

    Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.

    • Terminated Z-1898-2020 FDA Class II initiated 13 Feb 2020

    Post-implantation separation of an actuator end cap component that may expose internal components of the actuator, which could lead to hastened degeneration of the internal components and egress of Titanium alloy wear debris and resultant localized tissue discoloration.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
PGN
Device Class
Class II
Regulation
21 CFR Part 888.3070
Advisory Committee
Orthopedic
FDA Decision
SESE Substantially Equivalent