Skip to content

MAGNETOM Prisma 10849583

by Siemens Healthcare GmbH

MAGNETOM Prisma 10849583 is a full-body MRI system (FDA Class II) made by Siemens Healthcare GmbH, cleared under FDA 510(k) K231560 on 23 Oct 2023 (2 earlier submissions) and listed in FDA GUDID since 24 Sep 2016 (UDI-DI 04056869006734), active as of 12 Nov 2026. 6 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

6 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source GUDID, synced 12 Nov 2026.

Regulatory status

as of registry sync 12 Nov 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK231560 Date23 Oct 2023DecidedStatusActive
United StatesGUDID listing Number04056869006734 UDI-DI Date24 Sep 2016ListedStatusListed GUDID: in commercial distribution

Official records

From the FDA record

Device description
MRI Scanner
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 12 Nov 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K231560 decided 23 Oct 2023 Active· +2 more in the status table
  • FDA GUDID UDI-DI 04056869006734 listed 24 Sep 2016 Listed

Recalls

6 recalls · 6 ongoing · FDA

  • 5 recalls, same reason

    • Ongoing Z-2658-2025 FDA Class I initiated 28 Aug 2025
    • Ongoing Z-2659-2025 FDA Class I initiated 28 Aug 2025
    • Ongoing Z-2660-2025 FDA Class I initiated 28 Aug 2025
    • Ongoing Z-2661-2025 FDA Class I initiated 28 Aug 2025
    • Ongoing Z-2667-2025 FDA Class I initiated 28 Aug 2025

    There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

    • Ongoing Z-2648-2023 FDA Class II initiated 1 Sep 2023 UDI-DI 04056869006734 listed

    In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential artifacts depend on the position of the headband, the amount of lubricating grease, as well as the sequence parameters used

Timeline

10 dated events

  1. Cleared FDA 510(k) K153343
  2. Listed FDA GUDID 04056869006734
  3. Cleared FDA 510(k) K173592
  4. Recall initiated FDA recalls Z-2648-2023
  5. Cleared FDA 510(k) K231560
  6. Recall initiated FDA recalls Z-2658-2025
  7. Recall initiated FDA recalls Z-2659-2025
  8. Recall initiated FDA recalls Z-2660-2025
  9. Recall initiated FDA recalls Z-2661-2025
  10. Recall initiated FDA recalls Z-2667-2025

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
MAGNETOM Prisma
Generic name
Full-body MRI system, superconducting magnet
Model / reference
10849583
Primary DI / UDI-DI
04056869006734
FDA product code
LNH, MOS, LNI
GMDN term
Full-body MRI system, superconducting magnet
Device class
Class II
Regulation number
892.1000
GUDID distribution status
In Commercial Distribution

Similar devices in this category

Other devices filed under Full-body MRI system, superconducting magnet.

Other devices from Siemens Healthcare GmbH

See all devices from Siemens Healthcare GmbH →

Properties

  • Serial number
Other properties (14, all answered No)