Skip to content

Magnetom Spectra

by Siemens Medical Solutions USA, Inc. 2 variants

510(k) K121160

Magnetom Spectra is a medical device (FDA Class II; FDA product code LNH, System, Nuclear Magnetic Resonance Imaging) made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K121160 on 16 Jul 2012, active as of 22 Oct 2026. 3 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK121160 Date16 Jul 2012ClearedStatusActive

Official records

From the FDA record

Received by the FDA
16 Apr 2012
FDA decision
16 Jul 2012 (Substantially Equivalent)
FDA review time
91 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K121160 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

3 recalls · 1 ongoing · FDA

    • Ongoing Z-2663-2025 FDA Class I initiated 28 Aug 2025

    There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

    • Terminated Z-2344-2019 FDA Class II initiated 9 Jul 2019

    Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.

    • Terminated Z-2464-2012 FDA Class II initiated 12 Sep 2012

    When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the head-foot direction, the coil can heat up in the area of the electronic housing (white plastic lids).

Timeline

4 dated events

  1. Cleared FDA 510(k) K121160
  2. Recall initiated FDA recalls Z-2464-2012
  3. Recall initiated FDA recalls Z-2344-2019
  4. Recall initiated FDA recalls Z-2663-2025

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
LNH
Device class
Class II
Regulation
21 CFR Part 892.1000
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code LNH (System, Nuclear Magnetic Resonance Imaging).

Other devices from Siemens Medical Solutions USA, Inc.

See all devices from Siemens Medical Solutions USA, Inc. →