# Magnetom Verio

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/magnetom-verio-lnh>
- Manufacturer: [Siemens Medical Solutions USA, Inc.](https://devicerts.com/manufacturers/siemens-medical-solutions-usa-inc) (United States)
- Name in the registry: MAGNETOM VERIO
- Device class: Class II (US FDA)
- FDA product code: LNH

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K072237 · decision 2007-10-10 · decision SESE · status: Active

## Recalls

- Z-2664-2025 · Class I · ongoing · initiated 2025-08-28: There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- Z-2464-2012 · Class II · terminated · initiated 2012-09-12: When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the head-foot direction, the coil can heat up in the area of the electronic housing (white plastic lids).
- Z-1364-2012 ·  · terminated · initiated 2012-03-26: During product monitoring, Siemens has discovered that the gradient cable connections on a few systems did not meet their specifications after installation. The cable connections overheated, leading to the emission of smoke and possible material damage.
- Z-1401-2011 ·  · terminated · initiated 2011-01-10: On some systems the gradient cable connections did not meet firm's specifications after installation

## Identifiers

- 510(k) Number: K072237 [FDA 510(k)]
- FDA product code: LNH [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 892.1000 [FDA 510(k)]
- Advisory committee: Radiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Magnetom Verio, Siemens Medical Solutions USA, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/magnetom-verio-lnh (data as of 2026-11-20).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-20
- JSON: <https://devicerts.com/devices/magnetom-verio-lnh.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/magnetom-verio-lnh
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.
