Mammomat Revelation
by Siemens Medi Cal Solutions, Inc.
510(k) K173408
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK173408 | Date21 Mar 2018 | StatusActive |
Recalls
-
- Terminated Z-2388-2020
If an examination is interrupted when the InSpect Projection View (PV) is active and the vacuum biopsy system is being used, the tube arm's manual movement range is decreased from +/-15 deg. to +/- 6 deg. This limitation may lead to difficulties in removing the vacuum biopsy system.
-
- Terminated Z-0160-2020
An error can occur when performing a biopsy using the InSpect function. The acquisition workstation may become unresponsive to normal user interaction, can only take place in rare cases
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MUE
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1715
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent