Maximus System
by Hill-Rom Services Pte. Ltd. 2 variants
510(k) K192143
Maximus System is a medical device (FDA Class II; FDA product code NHJ, Device, Positive Pressure Breathing, Intermittent) made by Hill-Rom Services Pte. Ltd., cleared under FDA 510(k) K192143 on 14 Feb 2020, active as of 22 Oct 2026. 7 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
7 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK192143 | Date14 Feb 2020 | StatusActive |
Official records
- FDA 510(k) database K192143
From the FDA record
- Received by the FDA
- 8 Aug 2019
- FDA decision
- 14 Feb 2020 (Substantially Equivalent)
- FDA review time
- 190 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-2551-2026
- Ongoing Z-2549-2026
- Ongoing Z-2550-2026
- Ongoing Z-2548-2026
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
-
- Ongoing Z-1637-2026
Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment. This correction is providing the home caregiver specific instructions for use.
-
- Ongoing Z-2321-2024
- Ongoing Z-2322-2024
Potential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.
Timeline
- Cleared FDA 510(k) K192143
- Recall initiated FDA recalls Z-2321-2024
- Recall initiated FDA recalls Z-2322-2024
- Recall initiated FDA recalls Z-1637-2026
- Recall initiated FDA recalls Z-2551-2026
- Recall initiated FDA recalls Z-2549-2026
- Recall initiated FDA recalls Z-2550-2026
- Recall initiated FDA recalls Z-2548-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- NHJ
- FDA applicant
- Hill-Rom Services Pte, Ltd.
- Device class
- Class II
- Regulation
- 21 CFR Part 868.5905
- Advisory committee
- Anesthesiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code NHJ (Device, Positive Pressure Breathing, Intermittent).
- Piper IPPB Model 2055 by Vortran Medical Technology 1 Inc.
- EMERSON COUGH ASSIST MODEL CA-3000 W/AUTOMATIC TIMING AND MODEL CM-3000 W/MANUAL TIMING by J. H. Emerson Co.
- CoughAssist 70 Series, Water Trap 1098720 by Respironics Inc.
- CoughAssist 70 Series, 6ft clear tube 1116178 by Respironics Inc.
- VITAL COUGH by Lung Assist, Inc.
- BIRD (IPV) NONCONTINUOUS VENTILATORS by Percussionaire Corporation
- KT Airway Clearance Device II (KT-3) by Hunan Ruijiong Biotech Co., Ltd.
- BiWaze Clear System by Abmrc LLC
- PERCUSSIVENEB, MODEL PN-2001 by Vortran Medical Technology 1 Inc.
- PEGASO A-COUGH PERC PGP-00070-230 by Dima Italia S.r.l.