  "format": "devicerts.device.v1",
  "cite_as": "Maximus System, Hill-Rom Services Pte. Ltd. deviCERTS medical device catalogue. https://devicerts.com/devices/maximus-system-nhj (data as of 2026-11-17).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/maximus-system-nhj",
    "json_url": "https://devicerts.com/devices/maximus-system-nhj.json",
    "markdown_url": "https://devicerts.com/devices/maximus-system-nhj.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-17",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/maximus-system-nhj",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Maximus System",
    "registry_name": "Maximus System",
    "url": "https://devicerts.com/devices/maximus-system-nhj",
    "manufacturer": {
      "name": "Hill-Rom Services Pte. Ltd.",
      "url": "https://devicerts.com/manufacturers/hill-rom-services-pte-ltd",
      "country": "Singapore",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K192143",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "NHJ"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K192143",
      "number_label": null,
      "device_udi_di": null,
      "date": "2020-02-14",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2551-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-05-21",
      "reason": "There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits.  The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.",
      "applies": "this device",
      "via": "K192143"
    },
    {
      "number": "Z-2549-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-05-21",
      "reason": "There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits.  The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.",
      "applies": "this device",
      "via": "K192143"
    },
    {
      "number": "Z-2550-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-05-21",
      "reason": "There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits.  The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.",
      "applies": "this device",
      "via": "K192143"
    },
    {
      "number": "Z-2548-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-05-21",
      "reason": "There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits.  The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.",
      "applies": "this device",
      "via": "K192143"
    },
    {
      "number": "Z-1637-2026",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2026-02-18",
      "reason": "Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment.  This correction is providing the home caregiver specific instructions for use.",
      "applies": "this device",
      "via": "K192143"
    },
    {
      "number": "Z-2321-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-05-30",
      "reason": "Potential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter.  The handset plug is required to ensure proper operation and ventilator gas flow.",
      "applies": "this device",
      "via": "K192143"
    },
    {
      "number": "Z-2322-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-05-30",
      "reason": "Potential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter.  The handset plug is required to ensure proper operation and ventilator gas flow.",
      "applies": "this device",
      "via": "K192143"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K192143",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "NHJ",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 868.5905",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Anesthesiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}