Skip to content

Medex MX20838

by ICU Medical, Inc. 2 variants

Medex MX20838 is a Class II device made by ICU Medical, Inc., cleared under FDA 510(k) K942377 on 16 Aug 1994 (2 earlier submissions) and listed in FDA GUDID since 19 Oct 2016 (UDI-DI 10351688508805), active as of 23 Oct 2026. 3 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

3 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number10351688508805 UDI-DI Date19 Oct 2016ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK942377 Date16 Aug 1994DecidedStatusActive

Official records

Across registries

2 registries list this device

  • FDA 510(k) K942377 decided 16 Aug 1994 Active· +2 more in the status table
  • FDA GUDID UDI-DI 10351688508805 listed 19 Oct 2016 Listed

Recalls

4 recalls · 3 ongoing · FDA

  • 3 recalls, same reason

    • Ongoing Z-0185-2025 FDA Class II initiated 24 Sep 2024
    • Ongoing Z-0184-2025 FDA Class II initiated 24 Sep 2024
    • Ongoing Z-0186-2025 FDA Class II initiated 24 Sep 2024

    Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    • Terminated Z-1549-2008 FDA Unclassified (pre-2012) initiated 24 Mar 2008 on shared 510(k) K790314

    Misbranded/mispackaged device was distributed. Product labeled to contain 72 inch Standard Drip ( IV) Administration Sets, actually contained 60 inch (IV) administration sets with a' Y' site.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

8 dated events

  1. Cleared FDA 510(k) K790314
  2. Cleared FDA 510(k) K790455
  3. Cleared FDA 510(k) K942377
  4. Recall initiated FDA recalls Z-1549-2008
  5. Listed FDA GUDID 10351688508805
  6. Recall initiated FDA recalls Z-0185-2025
  7. Recall initiated FDA recalls Z-0184-2025
  8. Recall initiated FDA recalls Z-0186-2025

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Medex
Generic name
Invasive-pressure external transducer, single-use
Model / reference
MX20838
Primary DI / UDI-DI
10351688508805
FDA product code
FPB, DPT, KRA
GMDN term
Invasive-pressure external transducer, single-use
Device class
Class II
Regulation number
880.5440, 870.2120, 870.1210
GUDID distribution status
In Commercial Distribution
Sterilisation method
Ethylene Oxide

Similar devices in this category

Other devices filed under Invasive-pressure external transducer, single-use.

Other devices from ICU Medical, Inc.

See all devices from ICU Medical, Inc. →

Properties

  • Supplied sterile
  • Sterilise before use
  • Single use
  • Prescription use
  • Lot / batch number
  • Expiration date
Other properties (9, all answered No)