Medex MX20838
by ICU Medical, Inc. 2 variants
Medex MX20838 is a Class II device made by ICU Medical, Inc., cleared under FDA 510(k) K942377 on 16 Aug 1994 (2 earlier submissions) and listed in FDA GUDID since 19 Oct 2016 (UDI-DI 10351688508805), active as of 23 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10351688508805 | Date19 Oct 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK942377 | Date16 Aug 1994 | StatusActive |
Official records
- AccessGUDID record 10351688508805
- FDA 510(k) database K942377
- FDA 510(k) database K790455
- FDA 510(k) database K790314
Across registries
- FDA 510(k) K942377 decided 16 Aug 1994 Active
- FDA GUDID UDI-DI 10351688508805 listed 19 Oct 2016 Listed
Recalls
-
- Ongoing Z-0185-2025
- Ongoing Z-0184-2025
- Ongoing Z-0186-2025
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
-
- Terminated Z-1549-2008
Misbranded/mispackaged device was distributed. Product labeled to contain 72 inch Standard Drip ( IV) Administration Sets, actually contained 60 inch (IV) administration sets with a' Y' site.
Timeline
- Cleared FDA 510(k) K790314
- Cleared FDA 510(k) K790455
- Cleared FDA 510(k) K942377
- Recall initiated FDA recalls Z-1549-2008
- Listed FDA GUDID 10351688508805
- Recall initiated FDA recalls Z-0185-2025
- Recall initiated FDA recalls Z-0184-2025
- Recall initiated FDA recalls Z-0186-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Medex
- Generic name
- Invasive-pressure external transducer, single-use
- Model / reference
- MX20838
- Primary DI / UDI-DI
- 10351688508805
- FDA product code
- FPB, DPT, KRA
- GMDN term
- Invasive-pressure external transducer, single-use
- Device class
- Class II
- Regulation number
- 880.5440, 870.2120, 870.1210
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Ethylene Oxide
Similar devices in this category