Medex SM872M
Medex SM872M is an electric infusion pump administration set (FDA Class II) made by ICU Medical, Inc., cleared under FDA 510(k) K952160 on 21 Sep 1995 (2 earlier submissions) and listed in FDA GUDID since 3 Jan 2017 (UDI-DI 10351688500458), active as of 23 Oct 2026.
Active
Valid until
No expiry
FDA recalls
7 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10351688500458 | Date3 Jan 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK952160 | Date21 Sep 1995 | StatusActive |
Official records
- AccessGUDID record 10351688500458
- FDA 510(k) database K952160
- FDA 510(k) database K943897
- FDA 510(k) database K790314
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- Labelled as not made with natural rubber latex
- Sterility
- Supplied sterile; sterilise before use (Ethylene Oxide)
Across registries
- FDA 510(k) K952160 decided 21 Sep 1995 Active
- FDA GUDID UDI-DI 10351688500458 listed 3 Jan 2017 Listed
Recalls
-
- Ongoing Z-0743-2024
- Ongoing Z-0741-2024
- Ongoing Z-0742-2024
- Ongoing Z-0744-2024
- Ongoing Z-0740-2024
- Ongoing Z-0747-2024
- Ongoing Z-0746-2024
There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
-
- Terminated Z-1549-2008
Misbranded/mispackaged device was distributed. Product labeled to contain 72 inch Standard Drip ( IV) Administration Sets, actually contained 60 inch (IV) administration sets with a' Y' site.
Timeline
- Cleared FDA 510(k) K952160
- Recall initiated FDA recalls Z-1549-2008
- Listed FDA GUDID 10351688500458
- Recall initiated FDA recalls Z-0743-2024
- Recall initiated FDA recalls Z-0741-2024
- Recall initiated FDA recalls Z-0742-2024
- Recall initiated FDA recalls Z-0744-2024
- Recall initiated FDA recalls Z-0740-2024
- Recall initiated FDA recalls Z-0747-2024
- Recall initiated FDA recalls Z-0746-2024
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Medex
- Generic name
- Electric infusion pump administration set, single-use
- Model / reference
- SM872M
- Primary DI / UDI-DI
- 10351688500458
- FDA product code
- FPA, FPB
- GMDN term
- Electric infusion pump administration set, single-use
- Device class
- Class II
- Regulation number
- 880.5440
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Ethylene Oxide
Similar devices in this category
Other devices filed under Electric infusion pump administration set, single-use.