Skip to content

Medrad Intego Pet Infusion System, Source Administration Set, Patient Adminstratin Set Model Int Sys 100, Int Css, Int C

by Medrad, Inc.

510(k) K100798

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK100798 Date23 Jun 2010ClearedStatusActive

Recalls

3 recalls · 1 ongoing · FDA

    • Ongoing Z-0910-2023 FDA Class II initiated 30 Nov 2022

    Black specks may be present on the packaged needles

    • Terminated Z-0400-2017 FDA Class II initiated 30 Sep 2016

    The Source Administration Sets (SAS) used with the Medrad Intego PET Infusion System do not meet pressure testing specifications in an occlusion situation when the product is more than one year past manufacturing date.

    • Terminated Z-0627-2016 FDA Class II initiated 30 Nov 2015

    The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
FRN
Models / Variants
Device Class
Class II
Regulation
21 CFR Part 880.5725
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent

Other devices from Medrad, Inc.

See all devices from Medrad, Inc. →