MEDRAD® Avanta Fluid Management Injection System
by Bayer Medical Care Inc. 12 variants
Ref AVA 500 BTLH-M
Active
Valid until
Not listed
FDA recalls
1 ongoingDocuments
None linked in EUDAMED
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number0616258TFCN-0060I3B | Date5 Feb 2026 | StatusActive |
| United States | GUDID listing GUDID | Number00616258027203 | Date6 May 2024 | StatusListed |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-2296-2026
Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
-
- Terminated Z-0870-2013
The seal between the white polyethylene and the clear plastic may be missing on the Hand Controller Sheath package, resulting in a potential breach of the sterility of this accessory.
-
- Terminated Z-2866-2011
Event was initially intended to be a product enhancement of the MEDRAD Avanta Fluid Management Injection System. Upgrades were initiated in October, 2007. Observance of a possible increase in complaints and adverse event reporting resulted in the firm issuing "Important Product Use Reminder" to remind users of proper procedures for using the system and the need for diligence in reducing air embo
-
- Terminated Z-1199-2010
The product may be defective resulting in a reduction of the saline delivery rate and inadequate air purging.
-
- Terminated Z-0083-06
- Terminated Z-0082-06
stress crack on the stopcock
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade Name
- MEDRAD® Avanta Fluid Management Injection System EUDAMEDMEDRAD® Avanta Heat Maintainer, Bottle - Medium GUDID
- Generic Name
- Angiography contrast medium injection system, line-powered, mobile
- Model / Ref / Catalog
- AVA 500 BTLH-M EUDAMEDAVA 500 BTLH-L GUDID
- Primary DI / UDI-DI
- 00616258027203, 00616258022109
- Basic UDI-DI
- 0616258TFCN-0060I3B
- 510(k) Number
- FDA Product Code
- DXT
- EMDN Code
- A030105 Fluids heating systems
- GMDN Term
- Angiography contrast medium injection system, line-powered, mobile
- Device Class
- Class I EUDAMEDClass II GUDID
- Regulation Number
- 870.1650
- EU market of registration
- Netherlands
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | AVA 500 BTLH-M | AVA 500 BTLH-L |
| Device Class | Class I | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000007050
- Authorised representative
- Bayer Medical Care B.V. NL-AR-000000240
Other devices from Bayer Medical Care Inc.
Properties
Other properties (21)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No