Skip to content

Interstim Therapy System for Urinary Control

by Medtronic Neuromodulation

PMA P970004

Active United States

PMA

Valid until

No expiry

PMA approval is not time-limited

FDA recalls

1 ongoing

Recall on shared PMA P970004; other models may be affected

Documents

PMA P970004↗

FDA record

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA PMA, synced 6 Oct 2026. What this means

Regulatory status

as of registry sync 6 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA PMAAPPR Approved NumberP970004 Date29 Sep 1997ApprovedStatusActive

Recalls

5 recalls · 1 ongoing · FDA

    • Ongoing Z-0372-2025 FDA Class II initiated 18 Sep 2024 on shared PMA P970004

    The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.

    • Terminated Z-0998-2022 FDA Class II initiated 28 Feb 2022

    There is a software anomaly with the InterStim X Clinician software application with results in the data entered in the Patient Information fields not e retained and a " Data Lost" notification being displayed.

    • Terminated Z-1147-2015 FDA Class II initiated 9 Jan 2015 on shared PMA P970004

    Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.

    • Terminated Z-0037-2008 FDA Unclassified (pre-2012) initiated 25 Jun 2007

    A software issue may cause a lower than expected battery capacity estimate when evaluating an InterStim II Implantable Neurostimulator (model 3058) with an N'Vision Clinician Programmer(Model 8840) before implant .

    • Terminated Z-0040-2008 FDA Unclassified (pre-2012) initiated 25 Jun 2007

    Software issue may cause false low battery alert to be displayed when interrogating an Interstim Implantable Neurostimulator (Model 3023) with an N'Vision Clinician Programmer (Model 8840), or with an Inter Stim iCon Patient Programmer (Model 3037).

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA PMA Active

Specification

PMA

Trade Name
Medtronic Interstim Therapy System for Urinary Control
Generic Name
Stimulator, electrical, implantable, for incontinence
PMA Number
FDA Product Code
EZW
Device Class
Class III
Regulation
21 CFR Part 876.5270
Advisory Committee
Gastroenterology, Urology
FDA Decision
APPR Approved

Overview

AI-generated summary, 18 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

The Interstim Therapy System is an implantable electrical stimulator designed for managing urinary urge incontinence. Classified as a Stimulator, Electrical, Implantable, For Incontinence, it delivers sacral nerve stimulation to modulate bladder function in patients unresponsive or intolerant to conservative treatments. The system comprises modular components, including leads, an implantable pulse generator (IPG), and accessories, working together to provide targeted neuromodulation therapy.

This system is implanted surgically and requires sterile conditions during procedure and device handling. It is intended for single-use of the IPG and leads, though the console programmer and control magnet may be reused across patients. Components are MRI-unsafe and must be stored according to manufacturer instructions. Authorized under FDA PMA (P970004), the system adheres to regulatory classification 3 (876.5270) and is supplied with specific model numbers (e.g., 3886, 3080, 7424) for traceability and compliance.

Other devices from Medtronic Neuromodulation

Questions

AI-generated answers, 18 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

  • What specific components of the Interstim Therapy System are intended for single-use, and which can be reused across patients?

    The Interstim Therapy System includes single-use components like the implantable pulse generator (IPG) and leads, which are designed for one-time surgical implantation. In contrast, reusable components such as the console programmer and control magnet are intended for use with multiple patients, as they are not implanted and are meant for programming and activation purposes.

  • How should the Interstim Therapy System components be stored, and are there any special handling requirements?

    The Interstim Therapy System components must be stored according to the manufacturer’s instructions, though the data does not specify exact storage conditions. Given that it is an implantable device requiring sterile conditions during surgery and handling, it’s critical to follow Medtronic’s guidelines to maintain sterility and functionality. Additionally, the system is MRI-unsafe, so components must be stored away from MRI environments to prevent interference or damage.

  • What model numbers are associated with the Interstim Therapy System, and why are they important?

    The Interstim Therapy System includes specific model numbers such as 3886, 3080 (leads), 7424 (IPG), and others listed in the specifications. These model numbers are essential for traceability, regulatory compliance, and ensuring the correct components are used during implantation and programming. They help healthcare providers and manufacturers track the device’s history and verify its suitability for use.

  • Is the Interstim Therapy System intended for patients who have not responded to conservative treatments, and how does it work?

    Yes, the Interstim Therapy System is specifically indicated for patients with urinary urge incontinence who have failed or cannot tolerate more conservative treatments. It works by delivering sacral nerve stimulation, which modulates bladder function through electrical impulses delivered by implanted leads connected to an IPG. This targeted neuromodulation helps regulate bladder activity and reduce incontinence symptoms.

  • What regulatory pathway was followed for the approval of the Interstim Therapy System, and what classification does it hold?

    The Interstim Therapy System was approved through the FDA’s Premarket Approval (PMA) pathway under PMA number P970004, which is a rigorous process for high-risk devices like implantable stimulators. It is classified under FDA regulation 876.5270 as a Class III device, reflecting its higher risk and the need for extensive clinical validation before approval. This classification ensures the system meets stringent safety and effectiveness standards.