Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/Pacemaker
by Medtronic Physio-Control Corp.
510(k) K012274
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK012274 | Date5 Feb 2002 | StatusActive |
Recalls
-
- Terminated Z-1903-2010
A failure on the power supply assembly can result in either "No DC power" or "No DC or AC power." A failure of DC (battery) power can result in a delay of defibrillation therapy if no AC (line) power is available and the device will not operate.
-
- Terminated Z-1722-2011
A component failure on the AC power supply assembly results in "NO AC POWER" operation. A failure of AC power can result in a delay of defibrillation therapy if backup battery power is depleted.
-
- Terminated Z-0185-2012
When users do not disconnect the test plug and reconnect the QUIK-COMBO electrodes to the therapy cable, the test plug signal can be misinterpreted as a patient waveform.
-
- Terminated Z-2388-2008
Potential for the coin battery to drain prematurely causing the monitor clock time and date to be incorrect and the service light indicator to illuminate.
-
- Terminated Z-1050-2008
Inoperable Device - Potential for delay in therapy or prevention of defibrillation therapy due to corrosion of the Printed Circuit Board Assemblies (PCBA).
-
- Terminated Z-0925-2007
White Screen Lock up: LifePak 20 defibrillators with software version 48 or version 52 may experience an intermittent timing issue during the power on sequence. This may lock up the device resulting in the monitor screen going white, and may cause a potential delay or prevention of shock treatment.
-
- Terminated Z-0886-2007
Defibrillator internal battery cell may contain a short which could reduce battery capacity and cause the device to shut down without warning while operating on backup (DC) battery power.
-
- Terminated Z-1502-06
Devices with v38 system software do not display a ''LOW BATTERY: CONNECT TO AC POWER'' message when the monitor is on backup (DC) battery power and may shut down without warning.
-
- Terminated Z-0711-06
Therapy connectors may have connector sockets that do not meet design specs - short drill depth. Interruption of the connections may cause unexpected, intermittent or complete loss of pacing and defibrillation functions.
-
- Terminated Z-0295-2007
LIFEPAK 20 may lock-up when attempting to power-up on DC within 2 seconds after removing AC power.
-
- Terminated Z-0116-06
- Terminated Z-0117-06
Potential for voltage pins in connector of therapy cable to break.
-
- Terminated Z-0984-05
The defibrillator may not operate if the device is turned on during an AC Loss Alert.
-
- Terminated Z-0971-04
The DC Power Adapter may fail to charge LIFEPAK NiCd 2.4 AH batteries used with the LIFEPAK 12 defibrillator/monitor.
-
- Terminated Z-1107-04
Therapy cable doesn't meet firm's standards/may not maintain connections sufficient to support ECG monitoring, external pacing or defibrillation; ECG connector with higher than normal connection resistance resulting in intermittent ECG signal; therapy connector has potential for impact damage.
-
- Terminated Z-0938-04
Potential for display to go blank due to an electrical component that may fail due to mechanical shock of vibration of the defibrillator.
-
- Terminated Z-0294-04
Device unable to recognize the connection to the hard paddles accessory.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LDD
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.5300
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent