# Merit Hi-Flex(Tm)

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/merit-hi-flex-tm-fpk>
- Manufacturer: [Merit Medical Systems, Inc.](https://devicerts.com/manufacturers/merit-medical-systems-inc) (United States)
- Name in the registry: MERIT HI-FLEX(TM)
- Device class: Class II (US FDA)
- FDA product code: FPK

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K883718 · decision 1989-01-30 · decision SESE · status: Active

## Recalls

- Z-2293-2025 · Class II · ongoing · initiated 2025-06-05: High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
- Z-1319-2014 · Class II · terminated · initiated 2014-03-13 · matched through shared submission K883718, may concern other models: The products are labeled as sterile but were not sterilized.
- Z-0621-03 ·  · terminated · initiated 2003-02-13 · matched through shared submission K883718, may concern other models: A recent design change feature in the cap closure design for stopcocks in tubing convenience kits could preclude proper sterilization.
- Z-0620-03 ·  · terminated · initiated 2003-02-13 · matched through shared submission K883718, may concern other models: A recent design change feature in the cap closure design for stopcocks in tubing convenience kits could preclude proper sterilization.
- Z-0622-03 ·  · terminated · initiated 2003-02-13 · matched through shared submission K883718, may concern other models: A recent design change feature in the cap closure design for stopcocks in tubing convenience kits could preclude proper sterilization.

## Identifiers

- 510(k) Number: K883718 [FDA 510(k)]
- FDA product code: FPK [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 880.5440 [FDA 510(k)]
- Advisory committee: General Hospital [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Merit Hi-Flex(Tm), Merit Medical Systems, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/merit-hi-flex-tm-fpk (data as of 2026-11-23).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-11
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/merit-hi-flex-tm-fpk.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/merit-hi-flex-tm-fpk
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.