Micra™ MC1VR01US
Micra MC1VR01US is an intracardiac pacemaker (FDA Class III; FDA product code PNJ, Leadless pacemaker) made by Medtronic, Inc., approved under FDA PMA P150033 on 6 Apr 2016 and listed in FDA GUDID since 13 Jan 2018 (UDI-DI 00643169711761), active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00643169711761 | Date13 Jan 2018 | StatusListed |
| United States | FDA PMAAPPR | NumberP150033 | Date6 Apr 2016 | StatusActive |
Official records
- AccessGUDID record 00643169711761
- FDA PMA database P150033
From the FDA record
- Device description
- TPS MC1VR01US MICRA US SCALABLE
- MRI safety
- MR Conditional
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA PMA P150033 decided 6 Apr 2016 Active
- FDA GUDID UDI-DI 00643169711761 listed 13 Jan 2018 Listed
Recalls
-
- Ongoing Z-2128-2025
- Ongoing Z-2127-2025
A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no reset has actually occurred and there are no parameter changes or loss of diagnostics. Prior to this software update, if an erroneous SmartSync message was displayed, confusion regarding the message could have caused a clinician to consider an unnecessary system revision. Medtronic has received 39 reports of this display error with two instances of unnecessary device explant through 30 April 2025.
-
- Terminated Z-2492-2019
- Terminated Z-2493-2019
Medtronic is updating the Micra Instructions for Use (IFU) and the Micra Implant Procedure Tip Card to include specific information on the removal of the tether to release the Micra Pacemaker from the Micra Delivery System during implant.
Timeline
- Approved FDA PMA P150033
- Listed FDA GUDID 00643169711761
- Recall initiated FDA recalls Z-2492-2019
- Recall initiated FDA recalls Z-2493-2019
- Recall initiated FDA recalls Z-2128-2025
- Recall initiated FDA recalls Z-2127-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Trade name
- Micra™
- Generic name
- Intracardiac pacemaker
- Model / reference
- MC1VR01US
- Primary DI / UDI-DI
- 00643169711761
- PMA number
- FDA product code
- PNJ
- GMDN term
- Intracardiac pacemaker
- Device class
- Class III
- Regulation number
- 870.3610
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Conditional
- Size
- Length — 105.0 Centimeter