  "format": "devicerts.device.v1",
  "cite_as": "Micro Targeting Drive System, Fhc, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/micro-targeting-drive-system-haw (data as of 2026-11-24).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/micro-targeting-drive-system-haw",
    "json_url": "https://devicerts.com/devices/micro-targeting-drive-system-haw.json",
    "markdown_url": "https://devicerts.com/devices/micro-targeting-drive-system-haw.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-24",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/micro-targeting-drive-system-haw",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Micro Targeting Drive System",
    "registry_name": "MICRO TARGETING DRIVE SYSTEM",
    "url": "https://devicerts.com/devices/micro-targeting-drive-system-haw",
    "manufacturer": {
      "name": "Fhc, Inc.",
      "url": "https://devicerts.com/manufacturers/fhc-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K011775",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "HAW"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K011775",
      "number_label": null,
      "device_udi_di": null,
      "date": "2001-08-30",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0312-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-10-04",
      "reason": "A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2345-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2346-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2347-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2348-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2355-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2356-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-0895-05",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-03-14",
      "reason": "Potential for non-sterile pin to contaminate sterile field",
      "applies": "this device",
      "via": "K011775"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K011775",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "HAW",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 882.4560",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Neurology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}