# Micro Targeting Drive System

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/micro-targeting-drive-system-haw>
- Manufacturer: [Fhc, Inc.](https://devicerts.com/manufacturers/fhc-inc) (United States)
- Name in the registry: MICRO TARGETING DRIVE SYSTEM
- Device class: Class II (US FDA)
- FDA product code: HAW

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K011775 · decision 2001-08-30 · decision SESE · status: Active

## Recalls

- Z-0312-2025 · Class I · ongoing · initiated 2024-10-04: A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.
- Z-2345-2018 · Class II · terminated · initiated 2018-05-10: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
- Z-2346-2018 · Class II · terminated · initiated 2018-05-10: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
- Z-2347-2018 · Class II · terminated · initiated 2018-05-10: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
- Z-2348-2018 · Class II · terminated · initiated 2018-05-10: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
- Z-2355-2018 · Class II · terminated · initiated 2018-05-10: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
- Z-2356-2018 · Class II · terminated · initiated 2018-05-10: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
- Z-0895-05 ·  · terminated · initiated 2005-03-14: Potential for non-sterile pin to contaminate sterile field

## Identifiers

- 510(k) Number: K011775 [FDA 510(k)]
- FDA product code: HAW [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 882.4560 [FDA 510(k)]
- Advisory committee: Neurology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Micro Targeting Drive System, Fhc, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/micro-targeting-drive-system-haw (data as of 2026-11-24).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-24
- JSON: <https://devicerts.com/devices/micro-targeting-drive-system-haw.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/micro-targeting-drive-system-haw
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.