# Micromed BRAIN QUICK system

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/micromed-brain-quick-system-gwq>
- Manufacturer: [Micromed S.p.A.](https://devicerts.com/manufacturers/micromed-s-p-a) (Italy)
- Device class: Class II (US FDA)
- FDA product code: GWQ

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K171384 · decision 2018-02-01 · decision SESE · status: Active

## Recalls

- Z-2353-2026 · Class II · ongoing · initiated 2026-05-04: Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.

## Identifiers

- 510(k) Number: K171384 [FDA 510(k)]
- FDA product code: GWQ [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 882.1400 [FDA 510(k)]
- Advisory committee: Neurology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Micromed BRAIN QUICK system, Micromed S.p.A. deviCERTS medical device catalogue. https://devicerts.com/devices/micromed-brain-quick-system-gwq (data as of 2026-11-26).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-26
- JSON: <https://devicerts.com/devices/micromed-brain-quick-system-gwq.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/micromed-brain-quick-system-gwq
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.