  "format": "devicerts.device.v1",
  "cite_as": "microTargeting™ Procedure Kit KT-PPK-S-00366, Fhc, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/microtargeting-procedure-kit-kt-ppk-s (data as of 2026-11-24).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/microtargeting-procedure-kit-kt-ppk-s",
    "json_url": "https://devicerts.com/devices/microtargeting-procedure-kit-kt-ppk-s.json",
    "markdown_url": "https://devicerts.com/devices/microtargeting-procedure-kit-kt-ppk-s.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-24",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/microtargeting-procedure-kit-kt-ppk-s",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "microTargeting™ Procedure Kit KT-PPK-S-00366",
    "registry_name": "microTargeting™ Procedure Kit KT-PPK-S-00366",
    "url": "https://devicerts.com/devices/microtargeting-procedure-kit-kt-ppk-s",
    "manufacturer": {
      "name": "Fhc, Inc.",
      "url": "https://devicerts.com/manufacturers/fhc-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Catalogue number",
        "values": [
          {
            "value": "KT-PPK-S",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "00873263003664",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K011775",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K011992",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K071364",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "HAW",
      "GZL"
    ],
    "emdn": [],
    "gmdn": [
      {
        "code": "32556",
        "term": "Depth electrode"
      }
    ],
    "description": null,
    "noindex": true
  },
  "authorizations": [
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "00873263003664",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2016-09-17",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K071364",
      "number_label": null,
      "device_udi_di": null,
      "date": "2007-07-25",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K011775",
      "number_label": null,
      "device_udi_di": null,
      "date": "2001-08-30",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K011992",
      "number_label": null,
      "device_udi_di": null,
      "date": "2001-08-14",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0312-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2024-10-04",
      "reason": "A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2345-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2346-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2347-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2348-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2353-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011992"
    },
    {
      "number": "Z-2355-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2356-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011775"
    },
    {
      "number": "Z-2359-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-10",
      "reason": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "applies": "this device",
      "via": "K011992"
    },
    {
      "number": "Z-1143-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-12-28",
      "reason": "The raster tab for the GL4K version 1.4 software may assign incorrect depths for waveform recordings and if used to identify the implantation target for the DBS lead, the implant may be placed in the wrong location.",
      "applies": "this device",
      "via": "K071364"
    },
    {
      "number": "Z-0486-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-10-08",
      "reason": "Measuring fixture is incorrectly graduated.",
      "applies": "this device",
      "via": "K011992"
    },
    {
      "number": "Z-0179-06",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-10-31",
      "reason": "Power Assist microTargeting System may advance beyond intended target",
      "applies": "this device",
      "via": "K011992"
    },
    {
      "number": "Z-0180-06",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-10-31",
      "reason": "Power Assist microTargeting System may advance beyond intended target",
      "applies": "this device",
      "via": "K011992"
    },
    {
      "number": "Z-0895-05",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-03-14",
      "reason": "Potential for non-sterile pin to contaminate sterile field",
      "applies": "this device",
      "via": "K011775"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "microTargeting™ Procedure Kit",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Depth electrode",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Model / Ref / Catalog",
          "values": [
            {
              "value": "KT-PPK-S-00366",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "KT-PPK-S",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "00873263003664",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K011775",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K011992",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K071364",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "HAW",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "GZL",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Depth electrode",
              "code": "32556",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "882.4560",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "882.1330",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Clinical & Handling",
      "fields": [
        {
          "label": "MRI safety",
          "values": [
            {
              "value": "MR Unsafe",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}