Skip to content

Midas Rex Legend 12BA30DC

by Medtronic Powered Surgical Solutions

Midas Rex Legend 12BA30DC is an orthopaedic bur (FDA Class II; FDA product code HBB, Motor, Drill, Pneumatic) made by Medtronic Powered Surgical Solutions, cleared under FDA 510(k) K170312 on 15 May 2017 (1 earlier submission) and listed in FDA GUDID since 12 Mar 2022 (UDI-DI 00763000444846), active as of 23 Oct 2026. 8 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

9 closed

Documents

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00763000444846 UDI-DI Date12 Mar 2022ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK170312 Date15 May 2017DecidedStatusActive

Official records

From the FDA record

Device description
TOOL 12BA30DC LGND 12CM 3MM BALL DIA CRS
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K170312 decided 15 May 2017 Active· +1 more in the status table
  • FDA GUDID UDI-DI 00763000444846 listed 12 Mar 2022 Listed

Recalls

9 recalls · FDA

    • Terminated Z-1602-2016 FDA Class II initiated 8 Apr 2016

    Incorrect color coding and attachment reference were included on the labeling of 10BA40D dissecting tools with lot number H5237248.

    • Terminated Z-0321-2016 FDA Class II initiated 21 Oct 2015

    The wire pass hole is not present on the tools.

    • Terminated Z-0546-2010 FDA Unclassified (pre-2012) initiated 30 Nov 2009

    Component of product attachment may be outside established tolerance with potential for footed portion to break during normal use.

    • Terminated Z-2156-2008 FDA Unclassified (pre-2012) initiated 30 May 2008

    Cutting flute geometry of surgical dissecting tools may be outside of established tolerance. This may lead to reduced cutting performance, inability to perform desired dissection, increased risk of overheating, and potential need for revision surgery.

    • Terminated Z-1574-05 FDA Unclassified (pre-2012) initiated 1 Sep 2005 on shared 510(k) K020069

    Potential compromise of sterile packaging for disposable dissecting tools.

    • Terminated Z-1067-05 FDA Unclassified (pre-2012) initiated 29 Jun 2005

    Complaints of finger control component breaking off and falling into open surgical wound.

    • Terminated Z-0278-05 FDA Unclassified (pre-2012) initiated 11 Nov 2004

    Softer hose increases the propensity for outer exhaust hose to kink or become occluded as a result of clamping. Complaints of hose rupture/burst have been reported.

    • Terminated Z-0213-05 FDA Unclassified (pre-2012) initiated 3 Nov 2004

    Reduced plastic thickness of lubricant/diffuser cartidge can crack under pressure causing leakage of lubricant as visible mist (appearance of smoke).

    • Terminated Z-0419-03 FDA Unclassified (pre-2012) initiated 11 Nov 2002

    Lubricant leaking from motor after sterilization causing bleed-through sterilization wrap and/or excess lubricant in instrument case.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

12 dated events

  1. Recall initiated FDA recalls Z-0213-05
  2. Recall initiated FDA recalls Z-0278-05
  3. Recall initiated FDA recalls Z-1067-05
  4. Recall initiated FDA recalls Z-1574-05
  5. Recall initiated FDA recalls Z-2156-2008
  6. Recall initiated FDA recalls Z-0546-2010
  7. Recall initiated FDA recalls Z-0321-2016
  8. Recall initiated FDA recalls Z-1602-2016
  9. Cleared FDA 510(k) K170312
  10. Listed FDA GUDID 00763000444846

2 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Midas Rex Legend
Generic name
Orthopaedic bur, single-use
Model / reference
12BA30DC
Primary DI / UDI-DI
00763000444846
510(k) Number
FDA product code
HBB
GMDN term
Orthopaedic bur, single-use
Device class
Class II
Regulation number
882.4370
GUDID distribution status
In Commercial Distribution

Similar devices in this category

Other devices filed under Orthopaedic bur, single-use.

Other devices from Medtronic Powered Surgical Solutions

See all devices from Medtronic Powered Surgical Solutions →

Properties

  • Supplied sterile
  • Single use
  • Prescription use
  • Lot / batch number
Other properties (11, all answered No)