  "format": "devicerts.device.v1",
  "cite_as": "Midas Rex Legend 12BA30DC, Medtronic Powered Surgical Solutions. deviCERTS medical device catalogue. https://devicerts.com/devices/midas-rex-legend-12ba30dc (data as of 2026-11-18).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/midas-rex-legend-12ba30dc",
    "json_url": "https://devicerts.com/devices/midas-rex-legend-12ba30dc.json",
    "markdown_url": "https://devicerts.com/devices/midas-rex-legend-12ba30dc.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-23",
    "status_as_of": "2026-11-18",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/midas-rex-legend-12ba30dc",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Midas Rex Legend 12BA30DC",
    "registry_name": "MIDAS REX LEGEND 12BA30DC",
    "url": "https://devicerts.com/devices/midas-rex-legend-12ba30dc",
    "manufacturer": {
      "name": "Medtronic Powered Surgical Solutions",
      "url": "https://devicerts.com/manufacturers/medtronic-powered-surgical-solutions",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "00763000444846",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K020069",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K170312",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "HBB",
      "HBC"
    ],
    "emdn": [],
    "gmdn": [
      {
        "code": "45553",
        "term": "Orthopaedic bur, single-use"
      }
    ],
    "description": null,
    "noindex": true
  },
  "authorizations": [
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "00763000444846",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2022-03-12",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K170312",
      "number_label": null,
      "device_udi_di": null,
      "date": "2017-05-15",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K020069",
      "number_label": null,
      "device_udi_di": null,
      "date": "2002-03-18",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1602-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-04-08",
      "reason": "Incorrect color coding and attachment reference were included on the labeling of 10BA40D dissecting tools with lot number H5237248.",
      "applies": "this device",
      "via": "K020069"
    },
    {
      "number": "Z-0321-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-10-21",
      "reason": "The wire pass hole is not present on the tools.",
      "applies": "this device",
      "via": "K020069"
    },
    {
      "number": "Z-0546-2010",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-11-30",
      "reason": "Component of product attachment may be outside established tolerance with potential for footed portion to break during normal use.",
      "applies": "this device",
      "via": "K020069"
    },
    {
      "number": "Z-2156-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-05-30",
      "reason": "Cutting flute geometry of surgical dissecting tools may be outside of established tolerance.  This may lead to reduced cutting performance, inability to perform desired dissection, increased risk of overheating, and potential need for revision surgery.",
      "applies": "this device",
      "via": "K020069"
    },
    {
      "number": "Z-1574-05",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-09-01",
      "reason": "Potential compromise of sterile packaging for disposable dissecting tools.",
      "applies": "shared submission",
      "via": "K020069"
    },
    {
      "number": "Z-1067-05",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-06-29",
      "reason": "Complaints of finger control component breaking off and falling into open surgical wound.",
      "applies": "this device",
      "via": "K020069"
    },
    {
      "number": "Z-0278-05",
      "classification": "",
      "status": "terminated",
      "initiated": "2004-11-11",
      "reason": "Softer hose increases the propensity for outer exhaust hose to kink or become occluded as a result of clamping.  Complaints of hose rupture/burst have been reported.",
      "applies": "this device",
      "via": "K020069"
    },
    {
      "number": "Z-0213-05",
      "classification": "",
      "status": "terminated",
      "initiated": "2004-11-03",
      "reason": "Reduced plastic thickness of lubricant/diffuser cartidge can crack under pressure causing leakage of lubricant as visible mist (appearance of smoke).",
      "applies": "this device",
      "via": "K020069"
    },
    {
      "number": "Z-0419-03",
      "classification": "",
      "status": "terminated",
      "initiated": "2002-11-11",
      "reason": "Lubricant leaking from motor after sterilization causing bleed-through sterilization wrap and/or excess lubricant in instrument case.",
      "applies": "this device",
      "via": "K020069"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "MIDAS REX LEGEND",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Orthopaedic bur, single-use",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Model / reference",
          "values": [
            {
              "value": "12BA30DC",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "00763000444846",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K020069",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K170312",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "HBB",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Orthopaedic bur, single-use",
              "code": "45553",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "882.4370",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}