Skip to content

Midas Rex Microsaws 625R69-05-Tb

by Medtronic Powered Surgical Solutions

Ref 625R69-05-TB

Active European Union. Also listed in GUDID (US).

CE MDR

Valid until

Not listed

EUDAMED holds no validity date

FDA recalls

1 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source EUDAMED, GUDID, synced 23 Oct 2026. What this means

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number0763000B00013827Y Basic UDI-DIDate18 Sep 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number00763000464677 UDI-DIDate18 Mar 2022ListedStatusListed GUDID: in commercial distribution

Certificates

0 certificates · EUDAMED

EUDAMED lists no certificate for this device's Basic UDI-DI.

A certificate issued for a device group may exist without being linked here.

Recalls

3 recalls · 1 ongoing · FDA

    • Ongoing Z-0158-2022 FDA Class II initiated 31 Aug 2021

    There was an increase in complaints related to tip breaks and wobble/vibration.

    • Terminated Z-0961-2022 FDA Class II initiated 15 Dec 2021

    As a result of Medtronic's integrated power console (authentication number: 222ACBZX00018000) being used with software version: v2.7.3.0, it was discovered that there was a defect that resulted in a difference between the set number of rotations and the actual number of rotations in a specific mode when the product was used with our company's M5 handpiece.

    • Terminated Z-2085-2016 FDA Class II initiated 13 May 2016

    Medtronic Powered Surgical Solutions testing of the Sagittal Saw Rocker Arm Seal identified a potential for leakage of fluid into the motor during use.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States GUDID listing Listed

Specification

EUDAMED · GUDID

Trade Name
Midas Rex Microsaws
Generic Name
Orthopaedic bur, single-use
Device Name
Midas Rex Reciprocating Blade
Model / Reference
625R69-05-TB
Primary DI / UDI-DI
00763000464677
Basic UDI-DI
0763000B00013827Y
510(k) Number
FDA Product Code
ERL, HBE
EMDN Code
Z12100985 Instruments for motorised neurosurgery systems - consumables
GMDN Term
Orthopaedic bur, single-use
Device Class
Class IIa EUDAMED
Class II GUDID
Regulation Number
874.4250, 882.4310
Market of Registration
Netherlands
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 1 field differs

FieldEUDAMEDGUDID
Device ClassEUDAMED
Class IIa
GUDID
Class II

Other devices from Medtronic Powered Surgical Solutions

See all devices from Medtronic Powered Surgical Solutions →

Properties

  • Sterile
  • Single Use
  • Prescription Use
  • Lot / Batch Number

Other properties (20)

  • Implantable: No
  • Reusable: No
  • Contains Latex: No
  • Reprocessed: No
  • CMR Substance: No
  • Human-derived: No
  • Human Tissues: No
  • Animal Tissues: No
  • Endocrine Disruptor: No
  • Administering Medicine: No
  • Measuring Function: No
  • Incorporates Medicinal Product: No
  • Over-the-Counter: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Serial Number: No
  • Expiration Date: No
  • Manufacturing Date: No