Midas Rex Microsaws ES210
by Medtronic Powered Surgical Solutions
Ref ES210
Midas Rex Microsaws ES210 is a bone/joint surgical power tool motor (EU MDR Class IIa; FDA Class II) made by Medtronic Powered Surgical Solutions, registered in EUDAMED on 14 Sep 2026 (Basic UDI-DI 0763000B000137787), cleared under FDA 510(k) K081475 on 17 Oct 2008 and listed in FDA GUDID since 29 Dec 2021 (UDI-DI 00763000431075), active as of 24 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
1 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number0763000B000137787 00763000431075 | Date14 Sep 2026 | StatusActive |
| United States | GUDID listing GUDID | Number00763000431075 | Date29 Dec 2021 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK081475 | Date17 Oct 2008 | StatusActive |
Official records
- EUDAMED public site 0763000B000137787
- AccessGUDID record 00763000431075
- FDA 510(k) database K081475
From the FDA record
- Device description
- SAW ES210 MIDAS REX RECIP FNGR CNTL ROHS
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Ethylene Oxide)
Across registries
- EUDAMED Basic UDI-DI 0763000B000137787 registered 14 Sep 2026 Active
- FDA 510(k) K081475 decided 17 Oct 2008 Active
- FDA GUDID UDI-DI 00763000431075 listed 29 Dec 2021 Listed
Recalls
-
- Ongoing Z-0158-2022
There was an increase in complaints related to tip breaks and wobble/vibration.
-
- Terminated Z-0961-2022
As a result of Medtronic's integrated power console (authentication number: 222ACBZX00018000) being used with software version: v2.7.3.0, it was discovered that there was a defect that resulted in a difference between the set number of rotations and the actual number of rotations in a specific mode when the product was used with our company's M5 handpiece.
-
- Terminated Z-2085-2016
Medtronic Powered Surgical Solutions testing of the Sagittal Saw Rocker Arm Seal identified a potential for leakage of fluid into the motor during use.
Timeline
- Cleared FDA 510(k) K081475
- Recall initiated FDA recalls Z-2085-2016
- Recall initiated FDA recalls Z-0158-2022
- Recall initiated FDA recalls Z-0961-2022
- Listed FDA GUDID 00763000431075
- Registered EUDAMED 0763000B000137787
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- Midas Rex Microsaws
- Generic name
- Bone/joint surgical power tool motor, electric
- Device name
- Midas Rex Microsaw
- Model / reference
- ES210
- Primary DI / UDI-DI
- 00763000431075
- Basic UDI-DI
- 0763000B000137787
- 510(k) Number
- FDA product code
- ERL, HBE
- EMDN code
- Z12100902 Drills for neurosurgery
- GMDN term
- Bone/joint surgical power tool motor, electric
- Device class
- Class IIa EUDAMEDClass II GUDID
- Regulation number
- 874.4250, 882.4310
- EU registration type
- Standard device
- EU market of registration
- Netherlands
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Ethylene Oxide
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device class | Class IIa | Class II |
Similar devices in this category
Other devices filed under DRILLS FOR NEUROSURGERY.
- HDS-121 - HEKADRILL HANDPIECE - ELECTRIC by Zethon Ltd
- A-C15 - 150MM ANGLED ATTACHMENT (2.38MM) by Zethon Ltd
- SmartTwist MR Hand Drill NGS-HD-01 by Clearpoint Neuro, Inc.
- Primado2 P200-PER by NAKANISHI INC.
- EasyDrill Cranial Perforator Ø14 x Ø11 x 1,5mm DM0210FAA by Micromar Indústria e Comércio Ltda
- EasyDrill Cranial Perforator Ø9 x Ø6 x 3mm DM0008FAA by Micromar Indústria e Comércio Ltda
- ELAN 4 GA844 by Aesculap AG
- Nefmed Yüksek Devirli Elmas Uç (140mm /6mm) N140BA6D by Nefmed medikal ürünleri inşaat otomotiv dış ticaret A.Ş.