Skip to content

Midas Rex Microsaws Rasp-Cc-1

by Medtronic Powered Surgical Solutions EUDAMEDGUDID

Active European Union Open recall

Ref RASP-CC-1

Regulatory status

as of 29 Oct 2026

Recall ongoing. 1 ongoing FDA recall names this device. See recalls

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number0763000B00013827Y Basic UDI-DIDate18 Sep 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number00763000464813 UDI-DIDate18 Mar 2022ListedStatusListed GUDID: in commercial distribution

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States GUDID listing Listed

Certificates

0 certificates · EUDAMED

EUDAMED lists no certificate for this device's Basic UDI-DI.

A certificate issued for a device group may exist without being linked here.

Recalls

3 recalls · 1 ongoing · FDA

    • Ongoing Z-0158-2022 FDA Class II initiated 31 Aug 2021

    There was an increase in complaints related to tip breaks and wobble/vibration.

    • Terminated Z-0961-2022 FDA Class II initiated 15 Dec 2021

    As a result of Medtronic's integrated power console (authentication number: 222ACBZX00018000) being used with software version: v2.7.3.0, it was discovered that there was a defect that resulted in a difference between the set number of rotations and the actual number of rotations in a specific mode when the product was used with our company's M5 handpiece.

    • Terminated Z-2085-2016 FDA Class II initiated 13 May 2016

    Medtronic Powered Surgical Solutions testing of the Sagittal Saw Rocker Arm Seal identified a potential for leakage of fluid into the motor during use.

Specification

EUDAMED · GUDID

Trade Name
Midas Rex Microsaws
Generic Name
Orthopaedic bur, single-use
Device Name
Midas Rex Reciprocating Blade
Model / Reference
RASP-CC-1
Primary DI / UDI-DI
00763000464813
Basic UDI-DI
0763000B00013827Y
510(k) Number
FDA Product Code
ERL, HBE
EMDN Code
Z12100985 Instruments for motorised neurosurgery systems - consumables
GMDN Term
Orthopaedic bur, single-use
Device Class
Class IIa EUDAMED
Class II GUDID
Regulation Number
874.4250, 882.4310
Market of Registration
Netherlands
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 1 field differs

FieldEUDAMEDGUDID
Device ClassEUDAMED
Class IIa
GUDID
Class II

Other devices from Medtronic Powered Surgical Solutions

See all devices from Medtronic Powered Surgical Solutions →

Properties

  • Sterile
  • Single Use
  • Prescription Use
  • Lot / Batch Number

Other properties (20)

  • Implantable: No
  • Reusable: No
  • Contains Latex: No
  • Reprocessed: No
  • CMR Substance: No
  • Human-derived: No
  • Human Tissues: No
  • Animal Tissues: No
  • Endocrine Disruptor: No
  • Administering Medicine: No
  • Measuring Function: No
  • Incorporates Medicinal Product: No
  • Over-the-Counter: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Serial Number: No
  • Expiration Date: No
  • Manufacturing Date: No