MiniMed 530G System
PMA P120010
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA PMAAPPR | NumberP120010 | Date26 Sep 2013 | StatusActive |
Recalls
-
- Ongoing Z-1751-2026
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
-
- Terminated Z-2544-2014
Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Generic Name
- Automated insulin dosing, threshold suspend
- PMA Number
- FDA Product Code
- OZO
- Device Class
- Class III
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- APPR Approved
Overview
AI-generated summary, 18 Sep 2026
The MiniMed 530G System is an automated insulin dosing device with threshold suspend functionality, intended for the continuous delivery of basal insulin at user-selectable rates and administration of insulin boluses in user-selectable amounts for the management of diabetes mellitus in persons aged sixteen years and older requiring insulin. It also provides continuous monitoring and trending of glucose levels in the fluid under the skin.
The system can be programmed to automatically suspend insulin delivery when the sensor glucose value falls below a predefined threshold. It is supplied as a reusable system with disposable components, intended for use in home or clinical settings under medical supervision, and has received FDA PMA authorization (P120010).
Other devices from Medtronic, Inc.
Questions
AI-generated answers, 18 Sep 2026
What age group is the MiniMed 530G System intended for?
The MiniMed 530G System is intended for persons sixteen years of age and older who require insulin for the management of diabetes mellitus, as stated in the approval documentation.
Does the MiniMed 530G System automatically stop insulin delivery under certain conditions?
Yes, the system can be programmed to automatically suspend delivery of insulin when the sensor glucose value falls below a predefined threshold value, as specified in the approval order.
What type of authorization does the MiniMed 530G System have?
The MiniMed 530G System has received FDA PMA (Premarket Approval) authorization, with application number P120010, as indicated in the device data.