MiniMed 670G System
PMA P160017
Active
Valid until
No expiry
FDA recalls
12 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA PMAAPPR | NumberP160017 | Date28 Sep 2016 | StatusActive |
Recalls
-
- Ongoing Z-2981-2026
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
-
- Ongoing Z-1742-2026
- Ongoing Z-1739-2026
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
-
- Ongoing Z-1261-2025
- Ongoing Z-1266-2025
- Ongoing Z-1265-2025
All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
-
- Ongoing Z-0003-2025
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
-
- Ongoing Z-0193-2023
Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.
-
- Ongoing Z-0855-2022
New or replacement insulin pumps are not pre-programmed with basal rates or other verified settings (bolus wizard settings, sensor settings, etc.), which must be set up and saved on pumps prior to use. Patients who do not realize this will receive no basal insulin, which can lead to insulin under-delivery. Insulin under-delivery can cause severe hyperglycemia, which may lead to diabetic ketoacidosis. Current device labeling does not clearly identify that these insulin pumps are capable of functioning with the default basal setting of 0.0 units/hr for all 24 hours and that no error message will appear if the default basal setting remains active. Furthermore, the labeling for the affected devices does not highlight that saving the basal settings requires multiple steps and failure to complete all steps will not save the new basal settings, nor that the steps required to set and save a basal profile and other parameters which are used as part of the bolus calculator function are different.
-
- Ongoing Z-1771-2021
Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery.
-
- Ongoing Z-0516-2021
Missing solder battery connection, which could interrupt power of insulin pump.
-
- Ongoing Z-0956-2020
There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.
-
- Terminated Z-2604-2017
- Terminated Z-2605-2017
Medtronic MiniMed is recalling the MiniMed 600 series insulin pump because it may become temporarily stuck, and the keypad becomes unresponsive.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Generic name
- Automated insulin dosing device system, single hormonal control
- PMA number
- FDA product code
- OZP
- Device class
- Class III
- Advisory committee
- Clinical Chemistry
- FDA decision
- APPR Approved
Overview
AI-generated summary, 18 Sep 2026
The MiniMed 670G System is an automated insulin dosing device system for the management of Type 1 diabetes mellitus in persons fourteen years of age and older requiring insulin. It continuously monitors glucose levels in the fluid under the skin and automatically adjusts basal insulin delivery based on Continuous Glucose Monitor sensor glucose values using SmartGuard technology.
The system consists of the MiniMed 670G insulin pump, Guardian Link (3) Transmitter, Guardian Sensor (3), One-Press Serter, and Contour NEXT Link 2.4 Glucose Meter. The Guardian Sensor (3) is intended for single use for 7 days, while the Guardian Link (3) Transmitter is for single-patient multi-use. The system requires a prescription and is authorized under FDA PMA.
Other devices from Medtronic Minimed, Inc.
Questions
AI-generated answers, 18 Sep 2026
What age group is the MiniMed 670G System approved for use in managing Type 1 diabetes?
The MiniMed 670G System is authorized for use in individuals fourteen years of age and older who require insulin for managing Type 1 diabetes mellitus. This approval is based on the device’s design for automated insulin dosing and continuous glucose monitoring in this specific age group.
How long can the Guardian Sensor (3) be used continuously before needing replacement?
The Guardian Sensor (3) is intended for single-use only and is designed for 7 days of continuous glucose monitoring. After this period, it should be replaced to ensure accurate and reliable data for the MiniMed 670G System’s automated insulin dosing functionality.
Is the Guardian Link (3) Transmitter reusable within the same patient, or is it single-use?
The Guardian Link (3) Transmitter is intended for single-patient multi-use, meaning it can remain with the same patient for repeated monitoring sessions as long as it remains functional and properly maintained. However, it is not designed for use across multiple patients.
What is the role of the Contour NEXT Link 2.4 Glucose Meter in this system, and how does it differ from the Guardian Sensor (3)?
The Contour NEXT Link 2.4 Glucose Meter is used for direct blood glucose measurements via fingertip or palm samples, providing quantitative results (20–600 mg/dL) to confirm trends observed by the Guardian Sensor (3). Unlike the sensor, which provides continuous subcutaneous glucose monitoring and trends (not intended for therapy adjustments), the meter delivers real-time blood glucose values for clinical decision-making. Therapy adjustments should always be based on the meter’s readings, not the sensor alone.
What components of the MiniMed 670G System require a prescription, and which can be purchased over the counter?
The entire MiniMed 670G System—including the insulin pump, Guardian Sensor (3), Guardian Link (3) Transmitter, and One-Press Serter—requires a prescription as it is a medical device for automated insulin dosing. However, the Contour NEXT Link 2.4 Glucose Meter and its test strips are over-the-counter (OTC) devices, available without a prescription for self-testing blood glucose levels.