MiniMed 780G Insulin Pump MMT-1894MD
Listed
Valid until
Not applicable
FDA recalls
6 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00199150047055 | Date8 Nov 2025 | StatusListed |
Recalls
-
- Ongoing Z-2982-2026
- Ongoing Z-2980-2026
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
-
- Ongoing Z-2698-2026
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
-
- Ongoing Z-1741-2026
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
-
- Ongoing Z-1635-2026
Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or hyperglycemia due to inaccuracies in insulin delivery (either under-delivery, over-delivery, or cessation of delivery).
-
- Ongoing Z-1238-2026
A software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- MiniMed 780G Insulin Pump
- Generic Name
- Ambulatory insulin infusion pump, electronic, software-dosing, scalar algorithm
- Model / Reference
- MMT-1894MD
- Primary DI / UDI-DI
- 00199150047055
- 510(k) Number
- FDA Product Code
- NDC, QFG
- GMDN Term
- Ambulatory insulin infusion pump, electronic, software-dosing, scalar algorithm
- Device Class
- Class II
- Regulation Number
- 868.1890, 880.5730
- GUDID distribution status
- In Commercial Distribution
- MRI Safety
- MR Unsafe
Other devices from Medtronic Minimed, Inc.
Properties
Other properties (11)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No