MiniMed Flex MMT-8072
Listed
Valid until
Not applicable
FDA recalls
1 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00763000848156 | Date21 May 2026 | StatusListed |
Recalls
-
- Ongoing Z-0088-2027
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- MiniMed Flex
- Generic Name
- Ambulatory insulin infusion pump, electronic, software-dosing, scalar algorithm
- Model / Reference
- MMT-8072
- Primary DI / UDI-DI
- 00763000848156
- 510(k) Number
- FDA Product Code
- QFG
- GMDN Term
- Ambulatory insulin infusion pump, electronic, software-dosing, scalar algorithm
- Device Class
- Class II
- Regulation Number
- 880.5730
- GUDID distribution status
- In Commercial Distribution
- MRI Safety
- MR Unsafe
Other devices from Medtronic Minimed, Inc.
Properties
Other properties (13)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Prescription Use: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No