Skip to content

MiniMed Go App

by Medtronic Minimed

510(k) K253512Ref MMT-8600

Active European Union and United States. Also listed in GUDID (US).

CE MDR 510(k)

Valid until

Not listed

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source EUDAMED, FDA 510(k), GUDID, synced 23 Oct 2026. What this means

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number0763000B000327286 Basic UDI-DI 00199150038954UDI-DIDate6 Jul 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number00199150038954 UDI-DI Date14 Jun 2026ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE 510(k) Substantially Equivalent · Traditional NumberK253512 Date8 Jan 2026ClearedStatusActive

Certificates

0 certificates · EUDAMED

EUDAMED lists no certificate for this device's Basic UDI-DI.

A certificate issued for a device group may exist without being linked here.

Recalls

2 recalls · 2 ongoing · FDA

    • Ongoing Z-0265-2027 FDA Not yet classified initiated 2 Sep 2026

    A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia.

    • Ongoing Z-2778-2026 FDA Class II initiated 16 Jun 2026

    For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States FDA 510(k) Active

Specification

EUDAMED · 510(k) · GUDID

Trade Name
MiniMed Go
Generic Name
Percutaneous interstitial fluid glucose monitoring system, electrochemical
Device Name
Minimed Go App (OUS iOS)
Model / Reference
MMT-8600
Primary DI / UDI-DI
00199150038954 EUDAMED
00199150038978 GUDID
Basic UDI-DI
0763000B000327286
510(k) Number
FDA Product Code
NDC
EMDN Code
W020106019099 Various diabetes monitoring instruments - other
GMDN Term
Percutaneous interstitial fluid glucose monitoring system, electrochemical
Device Class
Class IIa EUDAMED
Class II 510(k)GUDID
Regulation
21 CFR Part 868.1890
Special Device Type
Software
EU market of registration
Netherlands
Advisory Committee
Anesthesiology
FDA Decision
SESE Substantially Equivalent
GUDID distribution status
In Commercial Distribution
MRI Safety
MR Unsafe

Truth per registry

3 sources · 2 fields differ

FieldEUDAMED510(k)GUDID
Primary DI / UDI-DIEUDAMED
00199150038954
510(k)
—
GUDID
00199150038978
Device ClassEUDAMED
Class IIa
510(k)
Class II
GUDID
Class II

Overview

AI-generated summary, 18 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

The MiniMed Go App is a Drug Dose Calculator designed to integrate real-time continuous glucose monitoring (CGM) data with insulin dosing calculations. It provides automated insulin delivery adjustments and dose recommendations for users with diabetes, facilitating precise glucose management through a mobile platform.

The app is intended for use with compatible Medtronic insulin pump systems and CGM devices, enabling seamless data synchronization for insulin dose tracking, calculations, and reporting. Supplied as a software application for mobile devices, it requires no sterilization and is not classified as a reusable or single-use medical device. Compatibility and regulatory authorizations are aligned with FDA 510(k) clearance and MDR compliance, adhering to the specified medical specialty of Anesthesiology for administrative classification.

Other devices from Medtronic Minimed, Inc.

See all devices from Medtronic Minimed, Inc. →

Questions

AI-generated answers, 18 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

  • What is the primary purpose of the MiniMed Go App in diabetes management?

    The MiniMed Go App is designed to automate insulin dosing calculations by integrating real-time continuous glucose monitoring (CGM) data with insulin pump systems. It provides personalized dose recommendations to help users with diabetes achieve more precise glucose control through a mobile platform, reducing manual calculations and improving treatment accuracy.

  • How is the MiniMed Go App supplied, and does it require sterilization?

    The MiniMed Go App is supplied exclusively as a software application for mobile devices—it is not a physical device and therefore does not require sterilization. Since it is a digital solution, it is neither reusable nor single-use, aligning with its classification as a non-physical medical software tool.

  • Is the MiniMed Go App compatible with all insulin pumps or only specific Medtronic systems?

    The MiniMed Go App is specifically designed to work with compatible Medtronic insulin pump systems and CGM devices. It is not intended for use with non-Medtronic pumps or third-party systems, as seamless data synchronization is a core feature of its functionality.

  • What regulatory pathways does the MiniMed Go App follow, and how does this affect its use?

    The MiniMed Go App has received FDA 510(k) clearance and complies with the EU Medical Device Regulation (MDR). These approvals confirm its safety and performance for automated insulin dose calculations when used with authorized Medtronic hardware. Clinicians and users should ensure their specific pump and CGM models are listed as compatible to maintain regulatory compliance.

  • Why is the MiniMed Go App classified under Anesthesiology in administrative records, even though it’s for diabetes management?

    The administrative classification under Anesthesiology (as noted in the advisory committee and product code) is based on the regulatory filing category (e.g., NDC product code routing) and does not reflect its clinical indication. The app is explicitly intended for diabetes management—specifically, automated insulin dosing—and its use is tied to Medtronic’s insulin pump and CGM systems, not anesthesia procedures.

Properties

  • Over-the-Counter
  • Lot / Batch Number

Other properties (22)

  • Sterile: No
  • Single Use: No
  • Implantable: No
  • Reusable: No
  • Contains Latex: No
  • Reprocessed: No
  • CMR Substance: No
  • Human-derived: No
  • Human Tissues: No
  • Animal Tissues: No
  • Endocrine Disruptor: No
  • Administering Medicine: No
  • Measuring Function: No
  • Incorporates Medicinal Product: No
  • Prescription Use: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Serial Number: No
  • Expiration Date: No
  • Manufacturing Date: No